The FDA announced on Jan. 11, 2021, that Ecolab Inc is recalling Model ORS-320 Slush + Warmer Disc Drapes due to potential slits in the sterile barrier. This defect poses a risk of infection to patients receiving warmed irrigation fluid or surgical slush during procedures.
How to tell if you have it
- Brand: Ecolab
- Product: Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes
- Lot code: D192619A
- Quantity: 132 cases (3168 drapes)
- Distribution: US Nationwide in AL, AZ, CA, CO, FL, HI, ID, IL, IN, KS, LA, MA, MD, MI, MO, MT, NJ, NY, OH, OK, OR, PA, TN, TX, UT, WA; Japan
The hazard
Slits were identified in a low quantity of the drapes from a single lot, potentially breaching the sterile barrier between fluid in the basin and the slush/warming equipment. This breach could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.
Context
GovInfoNow's records show Ecolab Inc has had 9 other recalls since Feb. 12, 2016. Other recent recalls include the July 30, 2020, Ecolab Hand Sanitizer Recall for discoloration and foreign substances, and the May 10, 2019, Ecolab QUIK-CARE Hand Sanitizer Recall involving wrong alcohol ingredients. Similar recent recalls of surgical products include the Dec. 16, 2025, Vortex Surgical device recall for compromised sterile pouches, the Jan. 28, 2026, Medline surgical drapes recall for fabric delamination, and the June 24, 2026, Covidien sutures recall for packaging tears.