Lower risk Medical device recall · FDA · Published January 14, 2021

Ortho-Clinical VITROS VersaTips Recall: Deformed Tips Cause Sample Errors

Ortho-Clinical recalls VITROS VersaTips from specific lots due to deformities that may cause up to 16% sample measurement errors in lab tests. Affected lots include VITROS 250, 350, ECi, 5,1 FS, 3600, 4600, 5600, and XT7600 models.

🇺🇸 FDA recall #Z-1136-2021Medical deviceOrtho-Clinical ›
Take action

Contact the company

Recalling firmOrtho-Clinical Diagnostics, Inc.100 Indigo Creek Dr · Rochester, NY 14626-5101
✎ GovInfoNow plain-language explainer

What happened, in plain English

Ortho-Clinical is recalling VITROS VersaTips used in lab testing because some versions have deformities that can cause sample measurement errors up to 16%. This affects lab equipment accuracy and could lead to incorrect test results if not fixed. Read more at govinfonow.com/recall/63423-ortho-clinical-vitros-versatips-recall-deformed-tips-cause-sample-errors.

Source Summary by GovInfoNow · govinfonow.com/recall/63423-ortho-clinical-vitros-versatips-recall-deformed-tips-cause-sample-errors · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1136-2021

Cite this page

“Ortho-Clinical VITROS VersaTips Recall: Deformed Tips Cause Sample Errors.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63423-ortho-clinical-vitros-versatips-recall-deformed-tips-cause-sample-errors. Based on FDA recall #Z-1136-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1136-2021%22&limit=1).

Why it’s dangerous

Deformed VersaTips can cause sample measurement errors of up to 16% in immunometric and competitive assays. This means lab results might be inaccurate, potentially leading to incorrect diagnoses or treatment decisions.

Who’s affected

Lab users with VITROS 250, 350, ECi, 5,1 FS, 3600, 4600, 5600, or XT7600 models who use the VersaTips for sample dispensing. The risk is highest for labs that rely on precise measurements for critical tests.

How to tell if you have this one

Check if your VITROS 250, 350, ECi, 5,1 FS, 3600, 4600, 5600, or XT7600 models have lot number 413420. The VersaTips are used to dispense samples for MicroSlide and MicroWell assays.

  • Brand
    Ortho-Clinical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Lot Number

    Look for lot number 413420 on the VersaTips packaging or label.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 38 states.

Alaska: nationwideAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: named in the noticeNHWashington: named in the noticeWAIdaho: named in the noticeIDMontana: named in the noticeMTNorth Dakota: named in the noticeNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: named in the noticeORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: named in the noticeAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: nationwideDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: named in the noticeHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Arizona, Arkansas, California, Colorado, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin.

Recall timeline

FDA terminated this recall 23 months after the firm started it.

1. Firm started the recall: Jan 14, 202112. FDA classified it Class II: Feb 19, 202123. Published in FDA’s Enforcement Report: Mar 3, 202134. FDA terminated the recall: Dec 27, 2022420212022
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1136-2021 (January 14, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1136-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA