Steris Corporation voluntarily recalled OT1000 Series Orthopedic Surgical Tables on Jan. 21, 2021, after the FDA announced a hazard where glued bonds may separate, causing unexpected leg movement during procedures. The agency issued recall number Z-1149-2021 for the Class II device.
How to tell if you have it
- Brand: Steris
- Product: OT1000 Series Orthopedic Surgical Tables
- Model Numbers: OT1000, OT1100, OT1200
- Quantity: 272 tables total (227 domestic, 45 foreign)
The hazard
The FDA stated that the glued bond between the beam and the leg spar assembly may separate over time. This separation could result in unexpected movement of the affected leg spar if it occurs during a patient procedure.
Context
GovInfoNow's records show Steris Corporation has 31 other recalls on record since June 15, 2017. Recent Steris recalls include a Sept. 23, 2026 issue with HexaVue Focus Integration System Software and an Aug. 31, 2026 recall of VERIFY Biological/Integrator Challenge Pack Model LCB006. Other recent recalls of similar products include IMRIS ORT200 Operating Room Table Recall (Aug. 6, 2026), IMRIS ORT300 Operating Room Table Recall (Aug. 6, 2026), and Think Surgical TMINI Tracking Arrays (May 8, 2026).