Biomerieux VITEK 2 Compact: False Test Results Risk with HL7 Connection

Biomerieux recalls VITEK 2 Compact (versions 8.01-9.02) with HL7 configured external communication due to potential false susceptible test results. Lab technicians may need to manually correct LIS results, risking incorrect treatment decisions.

🇺🇸 FDA recall #Z-1308-2021Medical deviceBiomerieux ›
Take action

Contact the company

Recalling firmBiomerieux Inc595 Anglum Rd · Hazelwood, MO 63042-2320
✎ GovInfoNow plain-language explainer

Biomerieux is recalling certain VITEK 2 Compact systems that may send false test results to laboratory information systems when connected via HL7. This could lead to incorrect treatment decisions if lab technicians have to manually adjust results. Affected units have specific version numbers and HL7 configuration. Read more at govinfonow.com/recall/63338-biomerieux-vitek-2-compact-false-test-results-risk-with-hl7.

Source Summary by GovInfoNow · govinfonow.com/recall/63338-biomerieux-vitek-2-compact-false-test-results-risk-with-hl7 · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1308-2021

Cite this page

“Biomerieux VITEK 2 Compact: False Test Results Risk with HL7 Connection.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63338-biomerieux-vitek-2-compact-false-test-results-risk-with-hl7. Based on FDA recall #Z-1308-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1308-2021%22&limit=1).

The FDA announced on Jan. 28, 2021, that Biomerieux Inc is recalling VITEK 2 Compact systems due to a potential safety risk of false susceptible test results. The agency stated the recall was voluntary and initiated by the firm.

How to tell if you have it

  • Brand: Biomerieux
  • Product: VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x))
  • Configuration: External communication configured to HL7
  • Units affected: 6777 systems

The hazard

BioMérieux identified a potential safety risk where the worst case is a false susceptible erroneous test result. The problem occurs when HL7 Connection is used, causing results in the VITEK 2 to not match those sent to the LIS, or preventing expert interpretation results and Therapeutic Corrections from being sent if an isolate is sent more than once.

Context

GovInfoNow's records show Biomerieux Inc has 36 other recalls on record since 2017-07-12. Other recent recalls of similar products include Biomerieux VITEK 2 AST-N461 Test Kit Recall (2026-07-24), Biomerieux VITEK 2 AST-N447 Kit Recall (2026-06-04), and Biomerieux VITEK 2 Gram-negative Susceptibility Cards May Give False Antibiotic Results (2026-01-05). Other recent recalls of similar products include HardyDisk AST Cefiderocol 30μg FDC30: False Susceptibility Results Risk (2026-06-08), BioFire Respiratory Panel 2.1 Recall: False Negatives Risk (2025-10-22), and Vivoo Vaginal pH Test Recall: False Results Risk Inappropriate Medical Treatment (2025-11-22).

How to tell if you have this one

Check if your VITEK 2 Compact has version numbers 8.01, 8.02 (8.0x), 9.01, or 9.02 (9.0x) with HL7 external communication configured. Each unit has a unique serial number and specific reference numbers.

  • Brand
    Biomerieux
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.
Recall numberProduct, codes & lotsQuantity
Z-1307-2021VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.; VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. (System: VITEK 2 60, Reference Number: 27225, Catalog Number: C20, UDI: 03573026162351); (System: VITEK 2 60, Reference Number: 27225R, Catalog Number: C20, UDI: 03573026206765); (System: VITEK 2 XL, Reference Number: 27227, Catalog Number: C20, UDI: 03573026162375); (System: VITEK 2 XL, Reference Number: 27227R, Catalog Number: C20, UDI: 03573026206789). Each instrument has unique serial number.8786 systems
Z-1308-2021VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.; VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. Each instrument has unique serial number. (System: VITEK 2 Compact 60, Reference Number: 27560, Catalog Number: C57, UDI: 03573026136796); (System: VITEK 2 Compact 30, Reference Number: 27560R, Catalog Number: C57, UDI: 03573026147501); (System: VITEK 2 Compact 30, Reference Number: 27530, Catalog Number: C57, UDI: 03573026136789); (System: VITEK 2 Compact 30, Reference Number: 27530R, Catalog Number: C57, UDI: 03573026147495); (System: VITEK 2 Compact 15, Reference Number: 27415, Catalog Number: C57, UDI: 03573026220624); (System: VITEK 2 Compact 15, Reference Number: 27415R, Catalog Number: C57, UDI: 03573026220631)6777 systems
Based on the official FDA remedy

What to do — step by step

  1. Check System Version

    Verify your VITEK 2 Compact's version number (8.01, 8.02, 9.01, or 9.02) and HL7 configuration.

How this recall unfolded

Recall timeline

FDA terminated this recall 23 months after the firm started it.

1. Firm started the recall: Jan 28, 202112. FDA classified it Class II: Mar 25, 202123. Published in FDA’s Enforcement Report: Mar 31, 202134. FDA terminated the recall: Jan 12, 2023420212023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1308-2021 (January 28, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1308-2021 to see it in context.