Getinge Group Logistics America, LLC recalled Flow i and Flow c anesthesia gas machines due to potential system check failures and incorrect carbon dioxide readings, the FDA said on Jan. 29, 2021.
How to tell if you have it
- Brand: Getinge
- Product: Flow i C20 Anesthesia System (UDI: 6677200 07325710001349)
- Product: Flow i C30 Anesthesia System (UDI: 6677300 07325710001349)
- Product: Flow i C40 Anesthesia System (UDI: 6677400 07325710001318)
- Product: Flow c Anesthesia System (UDI: 6887700 07325710009765)
- Units affected: 144
The hazard
Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.
Context
GovInfoNow's records show Getinge Group Logistics America, LLC has issued three other recalls since Sept. 17, 2020, including a Feb. 11, 2021 recall for hydraulic leakage in operating tables and a Dec. 21, 2020 recall for water leakage risks in washer disinfectors. Other recent recalls of similar products include the Aug. 7, 2026 Getinge Thoracic Catheter Recall and the June 30, 2026 Getinge SERVO-U/N Ventilator Sensor Module Recall.