Skip to content
Stop using now Medical device recall · FDA · Published January 29, 2021

Draeger ID Circuit Basic (N) 180 Breathing Circuits: Risk of Ventilation Leak During Use

Draeger Medical recalls ID Circuit Basic (N) 180 breathing circuits due to potential detachment causing ventilation leaks. All units are affected. Contact for replacement.

🇺🇸 FDA recall #Z-1180-2021Medical deviceDraeger Medical ›
Take action

Contact the company

Recalling firmDraeger Medical, Inc.3135 Quarry Rd · Telford, PA 18969-1042
✎ GovInfoNow plain-language explainer

Dra: Draeger Medical is recalling all ID Circuit Basic (N) 180 breathing circuits because a Y-piece may detach during use, causing leaks that could hinder patient ventilation. This is a serious safety issue for medical use. Read more at govinfonow.com/recall/63328-draeger-id-circuit-basic-n-180-breathing-circuits-risk-of.

Source Summary by GovInfoNow · govinfonow.com/recall/63328-draeger-id-circuit-basic-n-180-breathing-circuits-risk-of · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1180-2021

Cite this page

“Draeger ID Circuit Basic (N) 180 Breathing Circuits: Risk of Ventilation Leak During Use.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63328-draeger-id-circuit-basic-n-180-breathing-circuits-risk-of. Based on FDA recall #Z-1180-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1180-2021%22&limit=1).

The FDA announced on Jan. 29, 2021, that Draeger Medical, Inc. voluntarily recalled ID Circuit Basic (N) 180-Drager Breathing Circuits/Anesthesia Sets due to a potential leak hazard. The agency issued recall number Z-1180-2021 for the Class II device.

How to tell if you have it

  • Brand: Draeger Medical
  • Product: ID Circuit Basic (N) 180-Drager Breathing Circuits/Anesthesia Sets
  • Order Number: MP01327
  • Quantity: 25 units

The hazard

According to the FDA, the Y-piece may become detached from the ventilation hose either before or during the ventilation process. This detachment causes a leak that can hinder ventilation of the patient.

Context

GovInfoNow's records show Draeger Medical, Inc. has 16 other recalls on record since March 1, 2018. Recent similar product recalls include the Draeger ErgoStar CM airway connector issues reported on Nov. 10, 2025. Other recent recalls of similar products include the Draeger Atlan A300XL and A350XL Anesthesia Workstation recalls (Oct. 15, 2024) and the Draeger Perseus A5 battery shutdown risk recall (March 20, 2024).

How to tell if you have this one

Check for the ID Circuit Basic (N) 180 model with order number MP01327. All units are affected, so no specific lot or date is needed.

  • Brand
    Draeger Medical
  • Category
    Medical device
FDA recall event · 4 products

All 4 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    US Nationwide distribution.
Recall numberProduct, codes & lotsQuantity
Z-1177-2021VentStar Anesthesia (N) 180 - Drager Breathing Circuits/Anesthesia Sets Order Number: MP00333 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.; All lots4.450 units
Z-1178-2021VentStar Basic (N) 180- Drager Breathing Circuits/Anesthesia Sets Order Number: MP00353 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.; All lots1,525 units
Z-1179-2021VentStar Watertrap (N) 180-Drager Breathing Circuits/Anesthesia Sets Order Number: MP00363 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.; All lots150 units
Z-1180-2021ID Circuit Basic (N) 180-Drager Breathing Circuits/Anesthesia Sets Order Number: MP01327 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.; All lots25 units
Based on the official FDA remedy

What to do — step by step

  1. Stop Using Now

    Immediately stop using the ID Circuit Basic (N) 180 breathing circuits as they may detach and cause leaks.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA terminated this recall 9 months after the firm started it.

1. Firm started the recall: Jan 29, 202112. FDA classified it Class II: Mar 5, 202123. Published in FDA’s Enforcement Report: Mar 17, 202134. FDA terminated the recall: Nov 1, 202142021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1180-2021 (January 29, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1180-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA