The FDA announced on Feb. 11, 2021, that Getinge Group Logistics America, LLC is recalling its Operating Table Systems due to potential hydraulic leakage. This defect may cause unintended table movement, delaying surgical procedures.
How to tell if you have it
- Brand: Getinge
- Product: Operating Table System
- Part Numbers: 118001A0, 118001B0, 118001B1, 118001B2, 118001B3, 118001B4, 118001C0, 118001D0
- UDI codes: 04046768031500, 04046768031531, 04046768040700, 04046768088351, 04046768088061, 04046768124752, 04046768031555, 04046768031579
- Serial Numbers: 1 3402, 1 1028, 1 1264, 1 401, 3 74, 1 81, 2 1159, 2 - 762
- Units affected: 4288
The hazard
According to the FDA, hydraulic leakage at the inclination function may result in unintended Trendelenburg or anti-Trendelenburg movement of the table top. This issue depends on patient positioning and may lead to the patient having to be transferred to another OR table, causing a delay in the procedure.
Context
GovInfoNow's records show Getinge Group Logistics America, LLC has issued three other recalls since Sept. 17, 2020. On Jan. 29, 2021, the company recalled Flow i Anesthesia Machines due to FiCO2 errors, and on Dec. 21, 2020, it recalled CM320 Washer Disinfectors for water leakage risks. Other recent recalls of similar products include IMRIS ORT200 Operating Room Table Recall (Aug. 6, 2026), IMRIS ORT300 Operating Room Table Recall (Aug. 6, 2026), and Philips Azurion 7 B20 System Recall (Jan. 26, 2026).