Biosense Webster, Inc. recalled the CARTO VIZIGO Bi-Directional Guiding Sheath due to a potential hemostatic valve dislodgement hazard that could cause bleeding or embolism, according to the FDA on Feb. 26, 2021.
How to tell if you have it
- Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE
- UPC 10846835016260
The hazard
The hemostatic valve may dislodge when introducing a dilator or device into the sheath. This defect could result in minor bleeding and/or embolism.
Context
GovInfoNow's records show Biosense Webster, Inc. has had 7 other recalls on record since June 14, 2017. Recent brand recalls include the Jan. 5, 2025 VARIPULSE Catheter Recall for neurovascular event risk and the Feb. 16, 2024 CARTO VIZIGO Bi-Directional Guiding Sheath recall due to a manufacturing mix-up. Other recent recalls of similar products include the June 11, 2026 Boston Scientific FARADRIVE Steerable Sheath Recall for risk of air embolism and the June 11, 2026 Boston Scientific FARADRIVE Steerable Sheath Clear IDE Recall for risk of air embolism from valve damage.