The FDA announced on March 5, 2021, that Covidien is recalling its Surgiwand II Suction and Irrigation Device due to particulate matter hazards. The agency issued recall number Z-1411-2021 for the electrosurgical device.
How to tell if you have it
- Brand: Covidien
- Product: Surgiwand II Suction and Irrigation Device
- Product Number/CFN: 178083, 178093, 178094
- GTIN/UPN: 10884523000887, 20884523000884, 10884523000900, 20884523000907, 10884523000917, 20884523000914
- Units affected: 59,828 devices
The hazard
The FDA said investigations into customer complaints identified particulate matter inside the device tubing components. The root cause is a damaged Y-Connector component that generates debris during the assembly process connecting the tubing to the device.
Context
GovInfoNow's records show Covidien has had 14 other recalls on record since February 10, 2023. Recent examples include a June 23, 2025 recall of the Bedside SpO2 Monitor for alarm failures and a May 21, 2025 recall of the Newport HT70 Ventilator for shutdown risks. Other recent recalls of similar products include the Surgipro II S8698GX (June 24, 2026) and Dermalon 4-0 Blue 75cm Sutures (June 24, 2026), both citing infection risks from packaging issues.