Skip to content
Lower risk Medical device recall · FDA · Published March 9, 2021

Medtronic Percept PC Clinician Programmer Software Recalled: Cycling and Log Fill Issues

Medtronic recalls specific versions of its Percept PC clinician programmer software due to cycling and log fill problems that could cause crashes when copying settings or reading logs.

🇺🇸 FDA recall #Z-1368-2021Medical deviceMedtronic ›
Take action

Contact the company

Recalling firmMedtronic Neuromodulation7000 Central Ave Ne · Minneapolis, MN 55432-3568
✎ GovInfoNow plain-language explainer

Medtronic is recalling certain versions of its Percept PC clinician programmer software. These versions may cause crashes when copying settings from an older device or reading full system logs, potentially disrupting medical device programming. Read more at govinfonow.com/recall/63111-medtronic-percept-pc-clinician-programmer-software-recalled-cycling-and-log.

Source Summary by GovInfoNow · govinfonow.com/recall/63111-medtronic-percept-pc-clinician-programmer-software-recalled-cycling-and-log · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1368-2021

Cite this page

“Medtronic Percept PC Clinician Programmer Software Recalled: Cycling and Log Fill Issues.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63111-medtronic-percept-pc-clinician-programmer-software-recalled-cycling-and-log. Based on FDA recall #Z-1368-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1368-2021%22&limit=1).

The FDA announced on March 9, 2021, that Medtronic Neuromodulation is recalling the A610 Clinician Programmer Application for Deep Brain Stimulation due to software issues. The hazard involves improper copying of cycling features and application crashes when system event logs are full.

How to tell if you have it

  • Brand: Medtronic
  • Product: A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Guide Percept PC Model B35200 Neurostimulator Application
  • Versions: 2.0.4584, 2.0.4594, 2.0.4605, and 3.0.1057*
  • Units affected: 5332 downloads

The hazard

According to the FDA, there are two issues with this correction action. First, when copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator using the "replacement" function, the cycling feature is not copied properly. Second, the clinician programmer A610 Software application versions 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057* crash when attempting to read the last records in a full Percept PC INS System Event log.

Context

GovInfoNow's records show Medtronic Neuromodulation has 45 other recalls on record since August 9, 2017. Other recent recalls of similar products include Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation (May 13, 2026), Medtronic SynchroMed Flex Infusion Software Issue (April 8, 2026), and Medtronic Vanta CP App Software Recall (December 3, 2025). Other recent recalls of similar products include Abbott Medical Merlin PCS 365 (May 12, 2026), Philips Patient Information Center iX Software Freezes (July 31, 2026), and GE Medical Systems Portrait Compute V1.2 Software Recall (July 20, 2026).

How to tell if you have this one

The affected software versions are 2.0.4584, 2.0.4594, 2.0.4605, and 3.0.1057 (which is only approved for Europe). Check your software version in the Percept PC clinician programmer application.

  • Brand
    Medtronic
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Look for the version number in the Percept PC clinician programmer application to confirm if it's one of the affected versions.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Worldwide distribution - US Nationwide distribution in Puerto Rico and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 36 days after the firm started it.

1. Firm started the recall: Mar 9, 202112. FDA classified it Class II: Apr 7, 202123. Published in FDA’s Enforcement Report: Apr 14, 202132021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1368-2021 (March 9, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1368-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA