Skip to content
Lower risk Medical device recall · FDA · Published March 11, 2021

HydraVision Digital Imaging System: Incorrect Dose Display Risk

LIEBEL-FLARSHEIM recalls HydraVision Digital Imaging Systems with Product ID 705559 (65KW) and 705560 (80KW) due to potential overestimation of radiation dose. This may lead to unnecessary radiation exposure during medical procedures.

🇺🇸 FDA recall #Z-1992-2021Medical deviceLIEBEL-FLARSHEIM ›
Take action

Contact the company

Recalling firmLIEBEL-FLARSHEIM COMPANY LLC1034 S Brentwood Blvd · Saint Louis, MO 63117-1223
✎ GovInfoNow plain-language explainer

LIEBEL-FLARSHEIM is recalling HydraVision Digital Imaging Systems with specific product IDs because they may display a higher radiation dose than actually delivered. This could result in unnecessary radiation exposure for patients during medical imaging procedures. Read more at govinfonow.com/recall/63100-hydravision-digital-imaging-system-incorrect-dose-display-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/63100-hydravision-digital-imaging-system-incorrect-dose-display-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1992-2021

Cite this page

“HydraVision Digital Imaging System: Incorrect Dose Display Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63100-hydravision-digital-imaging-system-incorrect-dose-display-risk. Based on FDA recall #Z-1992-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1992-2021%22&limit=1).

The FDA announced on March 11, 2021, that Liebel-Flarsheim Company LLC is recalling the HydraVision Digital Imaging System because the displayed dosage is higher than the actual delivered dose. The agency issued this Class II recall under number Z-1992-2021.

How to tell if you have it

  • Brand: LIEBEL-FLARSHEIM
  • Product: HydraVision Digital Imaging System
  • Product ID/Reorder Number: 705559 (65KW)
  • Product ID/Reorder Number: 705560 (80KW)
  • Model/NDC: HUTSHVDIS

The hazard

The recall addresses a failure to comply with the performance standard for fluoroscopy systems, specifically that the displayed air kerma deviates from actual values by more than 35 percent. This results in the dosage displayed on the generator console and table monitor being higher than the actual delivered dose.

Context

GovInfoNow's records show Liebel-Flarsheim Company LLC has 1 other recall on record since 2026-07-31, specifically the Optiray Imaging Bulk Package-350 Recall for particulate matter hazard. Other recent recalls of similar products include Siemens ARTIS pheno X-ray System Recalled: Zero Dose Display Risk (2026-03-09), RayStation/RayPlan Software 11A Recalls: Radiation Dose Calculation Risk (2025-11-28), and Siemens ARTIS Pheno VE30A/VE40A: Risk of inaccurate 3D scans due to radiation exposure (2026-03-12).

How to tell if you have this one

Check for Product ID/Reorder Number 705559 (65KW) or 705560 (80KW) on the HydraVision Digital Imaging System. These models were sold with the model number HUTSHVDIS.

  • Brand
    LIEBEL-FLARSHEIM
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Product ID

    Identify the Product ID/Reorder Number on your HydraVision Digital Imaging System, which should be 705559 (65KW) or 705560 (80KW).

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide Distribution and in the country of Canada” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 3.5 years after the firm started it.

1. Firm started the recall: Mar 11, 202112. FDA classified it Class II: Jul 2, 202123. Published in FDA’s Enforcement Report: Jul 14, 202134. FDA terminated the recall: Sep 18, 2024420212024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1992-2021 (March 11, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1992-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA