The FDA announced on March 18, 2021, that Fresenius Medical Care Holdings, Inc. is recalling AquaBplus and AquaBplus B2 Water Purification Systems due to a potential heat disinfection error. The hazard stems from an installation guideline mistake regarding mains used with the AquaBplus 2500 osmosis system.
How to tell if you have it
- Brand: Fresenius
- Products: AquaBplus and AquaBplus B2 Water Purification Systems
- Fresenius P/N: 24-2500-0
- AquaBplus 3000 - Typ USA (G02040109-US)
- AquaBplus 2500 - Typ USA (G02040115-US)
- AquaBplus HF- Typ USA (G02040120FMC-US)
- First stage serial numbers: 5BPS0601, 6BPS0933, 7BPS1457, 7BPS1526, 7BPS1458, 7BPS1527, 8BPS1801, 8BPS1800, 8BPS1870, 8BPS1883, 8BPS1898, 8BPS1885, 8BPS1884, 8BPS1918, 7BPS14443, 8BPS1930, 8BPS1928, 8BPS1929, 8BPS1937, 8BPS1927, 8BPS2052, 8BPS2081, 8BPS2108, 8BPS2157, 9BPS2234, 7BPS1441, 9BPS2512
- Second stage (B2) serial numbers: 5BDS0194, 6BDS067, 7BDS0692, 7BDS0726, 7BDS0643, 7BDS0722, 8BDS0907, 8BDS0909, 8BDS0958, 8BDS0960, 8BDS0964, 8BDS0963, 8BDS0965, 8BDS0978, 7BDS0684, 8BDS0989, 8BDS0997, 8BDS0995, 8BDS0998, 8BDS0992, 8BDS1075, 8BDS1087, 8BDS1102, 8BDS1139, 9BDS1181, 7BDS0683, 9BDS1373
- HF option serial numbers: 5BFS0125, 6BFS0241, 7BFS0413, 7BFS0456, 7BFS0446, 7BFS0486, 8BFS0541, 8BFS0609, 8BFS0644, 8BFS0665, 8BFS0663, 8BFS0577, 8BFS0621, 8BFS0616, 7BFS0439, 8BFS0686, 8BFS0702, 8BFS0703, 8BFS0695, 8BFS0676, 8BFS0727, 8BFS0757, 8BFS0776, 8BFS0805, 9BFS0845, 7BFS0477, 9BFS0997
- Distribution: US Nationwide in FL, MA, MD, MI, MO, NC, NJ, NY, PA, PR, SC, TN, TX
The hazard
An error in the US Installation Guideline instructed that 3-ring mains may be installed with an AquaBplus 2500 device instead of 2-ring mains. The FDA stated that heat disinfection using a 3-ring main and fluid fly loops may not be as effective as a 2-ring main installation.
Context
GovInfoNow's records show Fresenius Medical Care Holdings, Inc. has 44 other recalls on record since April 14, 2014. Other recent recalls of similar products include Fresenius Bicarby Dialysate Leaks Risk (May 12, 2026), Fresenius 5008X Blood Tubing Set Recall (May 11, 2026), and Fresenius Medical Care Bicarby Dialysate Recall (May 7, 2026). Additional recent recalls include Fresenius Medical Care 2008T BlueStar Hemodialysis Machine Recall (June 30, 2026), Fresenius Medical Care CombiSet Hemodialysis Tubing Set Recall (July 29, 2026), and Fresenius Medical Care 5008X HD/HDF with CLiC Blood Tubing Set Recall (August 10, 2026).