The FDA announced on March 22, 2021, that Normand-Info S.A.S.U. is recalling the Normand Remisol Advance Data Manager software device due to insufficient instructions for programming and testing rules. This deficiency may lead to erroneous laboratory results. Owners should contact the company for remedy details.
How to tell if you have it
- Brand: Normand
- Product: Normand Remisol Advance Data Manager
- Catalog Numbers: B08688, B22517, B35058, B74811, B92487
- Lot Numbers: All Lots
- Software versions 1.5 to 1.9
The hazard
According to the FDA, the Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.
Context
GovInfoNow's records show Normand-Info S.A.S.U. has two other recalls on record since August 15, 2019. On April 30, 2020, the company recalled Normand Info Remisol Advance Software 1.12 due to a risk of incorrect blood test results. Other recent recalls of similar products include Beckman Coulter REMISOL Advance Software Issue: Incorrect Test Results Risk (April 29, 2026), Beckman Coulter AU5800 Chemistry Analyzer Recall: Delay in Results Risk (January 30, 2026), and Beckman Coulter DxC 500i Modules with Software 1.5 Risk: Erroneous Test Results (July 15, 2026).