Philips North America LLC recalled Philips SureSigns VM4, VM6 and VM8 patient monitors distributed in China due to heart rate meter accuracy failures in pediatric mode, according to the FDA on March 23, 2021.
How to tell if you have it
- Brand: Philips
- Product names: SureSigns VM4, SureSigns VM6, SureSigns VM8
- Model numbers: 863063, 863085, 863064, 863065, 863086, 863066, 863068, 863087, 863088
- Other model numbers: 453564021961, 453564021971, 53564022001, 453564022041, 53564022081, 453564024371, 453564024401, 453564024421, 453564024441, 453564024461, 453564024481
- Software version: SW A.02.63
- Units affected: 31773
- Distribution: China
The hazard
In pediatric mode, when the input signal rate is over 300 bpm, the indicated heart rate will be lower than this upper limit. The products fail to comply with Chinese Standard YY1079-2008 clauses 4.2.6 and 4.2.7.3.
Context
GovInfoNow's records show Philips North America LLC has 129 other recalls on record since July 12, 2017. Recent recalls include the Achieva 1.5T systems recalled on August 21, 2026, and the Spectral CT 7500 Scanner recalled on August 18, 2026. Other recent recalls of similar products include Boston Scientific LUX-Dx II Plus Model M312 (March 30, 2026), Boston Scientific LUX-DX II Arrythmia Detector Recall (March 30, 2026), and Boston Scientific CRT-P Pacemakers (March 19, 2026).