The Biomed Guys Alaris Infusion Pump Module 8100 Recall: Bezel Repair Part Cracking Risk
The Biomed Guys recalls Alaris Infusion Pump Module Model 8100 serviced with non-original bezel repair parts. Cracking may cause infusion errors. Affected units have specific serial numbers.
Use caution until it’s fixed
Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Contact the company
What happened, in plain English
The Biomed Guys is recalling Alaris Infusion Pump Module Model 8100 units that were serviced with non-original bezel repair parts. These parts may crack, leading to infusion errors like over-infusion or interruption of medication flow. This could cause serious harm to patients using the pump. Read more at govinfonow.com/recall/62952-the-biomed-guys-alaris-infusion-pump-module-8100-recall-bezel.
Source Summary by GovInfoNow · govinfonow.com/recall/62952-the-biomed-guys-alaris-infusion-pump-module-8100-recall-bezel · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1651-2021
Cite this page
“The Biomed Guys Alaris Infusion Pump Module 8100 Recall: Bezel Repair Part Cracking Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62952-the-biomed-guys-alaris-infusion-pump-module-8100-recall-bezel. Based on FDA recall #Z-1651-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1651-2021%22&limit=1).
Why it’s dangerous
The non-original bezel repair part used during service may crack or separate. This can cause the pump to over-infuse or under-infuse medication, or stop the infusion entirely. These errors could lead to serious harm to patients.
Who’s affected
Patients using the Alaris Infusion Pump Module Model 8100 that were serviced by The Biomed Guys with the TIPA-8100-4410 bezel repair part are most at risk. The recall affects specific units with serial numbers listed in the notice.
How to tell if you have this one
Check if your Alaris Infusion Pump Module Model 8100 was serviced by The Biomed Guys using the TIPA-8100-4410 repair part. Affected units have specific serial numbers listed in the recall notice.
- Brand
The Biomed Guys- Category
Medical device
What to do — step by step
Check Serial Numbers
Review the list of affected serial numbers provided in the recall notice to determine if your pump is included.
Contact The Biomed Guys
Reach out to The Biomed Guys directly to confirm if your pump was serviced with the impacted bezel repair part.
Where it went and when
Where it was distributed
Distributed in Indiana and Texas.
FDA distribution: “Us distribution to the states of IN and TX.” States read from FDA’s own wording; “not listed” means the notice does not name it.
States: Indiana, Texas.
Recall timeline
FDA published this recall 2 months after the firm started it.
- Firm started the recall
- FDA classified it Class I
- Published in FDA’s Enforcement Report
Dates from the agency record only; steps without a date are left out.
Similar recalls
Latest medical device recalls
Keep exploring
Where this comes from
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
- Official source: FDA Recall #Z-1651-2021 (March 25, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1651-2021 to see it in context.
- Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
- GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.