The FDA announced on March 26, 2021, that Infusion Pump Repair is recalling Alaris Pump Module Model 8100 units serviced with an impacted bezel repair part. The agency stated the recall was voluntary and initiated by the firm.
How to tell if you have it
- Brand: Infusion Pump Repair
- Product: Alaris Pump Module, Model 8100
- Distribution: California
- Sale Order/Serial Numbers: 1599930872/12335834, 13617863, 13447266, 13089105, 12333728, 13616451, 13618575, 12729374; 82943/4122252, 4122320; 01112101A/12638236, 13813590, 12639090, 13744738, 9951014
- Units affected: 15
The hazard
According to the FDA, bezel repair parts used to service and repair infusion pump modules may experience cracking or separation of the posts from the bezel over time. This separation may result in free flow, over infusion, under infusion, or interruption of infusion.
Context
GovInfoNow's records show Infusion Pump Repair has been involved in other recent recalls of similar products, including the Fresenius Kabi LVP-0004 Infusion Pump Recall (2026-02-04) and the B. Braun Medical Spinal Epidural Pump Model 530184 Recall (2026-01-29). Other recent recalls of similar products include the Baxter Novum IQ Syringe Pump Recall (2026-08-11).