The FDA announced on March 29, 2021, that Philips North America LLC voluntarily recalled Xper Flex Cardio monitoring devices due to performance issues including delayed parameter displays and software crashes. Owners should contact Philips for a remedy.
How to tell if you have it
- Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices
- Service numbers: 453564241901, 453564483321, 453564621791, 453564675021, 453564669081, 453564674581, 453564845841, 453564845881, 453564241911, 453564483331, 453564621801, 453564675001, 453564845861
- Software versions: 1.5.19 and 1.6.8
- UPCs: 00884838082113, 00884838083516, 00884838083820, 00884838084902, 00884838084919, 00884838086005, 00884838093645, 00884838093652, 00884838095052, 00884838103245
The hazard
According to the FDA, performance issues with the Xper Flex Cardio Physio Monitoring System include potential delays in displaying ECG and blood pressure parameters, patient weight rounding errors, periodic software crashes, and missing SpO2 displays. The record states that waveforms may freeze or become flat lines without audible alarms, and unexpected non-physiological data may appear upon startup.
Context
GovInfoNow's records show Philips North America LLC has 129 other recalls on file since July 12, 2017. Recent brand recalls include an August 21, 2026 recall of Achieva 1.5T systems for software termination issues and an August 18, 2026 recall of Spectral CT 7500 Scanners for potential extra radiation exposure. Other recent recalls of similar products include a July 31, 2026 recall of Philips PIC iX Essentials Software for patient data issues, an August 6, 2026 recall of Boston Scientific LUX-Dx Cardiac Monitor Systems for delayed alerts, and a March 30, 2026 recall of Boston Scientific LUX-DX II Arrhythmia Detectors for missing heart monitoring data.