The FDA announced on March 29, 2021, that Randox Laboratories Ltd. recalled Calibration Serum Level 3 (CAL 3) due to a positive bias in Total and Direct Bilirubin results. The agency issued recall number Z-1563-2021 for the Class II device.
How to tell if you have it
- Brand: Randox
- Product: Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator
- Catalogue Number: CAL2351
- Lot number: 1162UE
- UPC: 05055273200966
The hazard
The calibrator exhibits a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across assay ranges, attributed to bilirubin instability. This error is evident in both Quality Control and Patient samples.
Context
GovInfoNow's records show Randox Laboratories Ltd. has 49 other recalls on record since June 7, 2017. Recent brand recalls include a March 5, 2024 recall for Inorganic Phosphate Bias in Calibration Serum and a June 7, 2024 recall regarding missing FDA 510(k) clearance for Cystatin C Reagent. Other recent recalls of similar products include the Siemens Atellica CH Urine Albumin recall (July 6, 2026), the Siemens Dimension Tacrolimus Flex Reagent Cartridge recall (May 7, 2026), and the Beckman Coulter Access Total T4 Calibrator recall (March 16, 2026).