Skip to content
Lower risk Medical device recall · FDA · Published April 20, 2021

Siemens Multitom RAX Software VF10 Recall: Image Misalignment Risk

Siemens recalls Multitom RAX with software version VF10 due to potential image misalignment and dose calculation errors. Affected units may show incorrect imaging or dose warnings.

🇺🇸 FDA recall #Z-1685-2021Medical deviceSiemens ›
Take action

Contact the company

Recalling firmSiemens Medical Solutions USA, Inc40 Liberty Blvd · Malvern, PA 19355
✎ GovInfoNow plain-language explainer

Siemens is recalling Multitom RAX CT scanners with software version VF10 because of two issues: the system might misalign the X-ray beam by more than 20% during scans, and it could incorrectly calculate radiation doses, triggering warnings. These problems are unlikely to cause harm in normal use but could lead to inaccurate images or unnecessary radiation alerts. Read more at govinfonow.com/recall/62763-siemens-multitom-rax-software-vf10-recall-image-misalignment-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/62763-siemens-multitom-rax-software-vf10-recall-image-misalignment-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1685-2021

Cite this page

“Siemens Multitom RAX Software VF10 Recall: Image Misalignment Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62763-siemens-multitom-rax-software-vf10-recall-image-misalignment-risk. Based on FDA recall #Z-1685-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1685-2021%22&limit=1).

The FDA announced on April 20, 2021, that Siemens Medical Solutions USA, Inc. is recalling Multitom RAX units with software version VF10 due to potential radiation exposure errors. The agency mandated the recall of 26 devices distributed nationwide.

How to tell if you have it

  • Brand: Siemens
  • Product: Multitom RAX
  • Software version: VF10
  • Model number: 10860700

The hazard

Siemens stated that the system may allow image acquisition when more than 20% of the beam misses the detector, potentially causing overshoot. In very rare cases, the calculated dose value may exceed the limit of 2 Gy due to incorrect source-to-image distance calculations, causing the buzzer to activate erroneously.

Context

GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Recent similar recalls include the Siemens ARTIS icono Ceiling Monitor Recall (June 8, 2026), the Siemens ARTIS icono floor recall (May 20, 2026), and the Siemens ARTIS Pheno VE30A/VE40A recall (March 12, 2026). Other recent recalls of similar products include Philips Ingenuity CT Scanners (August 7, 2026), Philips Brilliance CT 64 Medical Scanner Recall (August 7, 2026), and Philips IQon Spectral CT Software 4.7.7.x Recall (August 7, 2026).

How to tell if you have this one

Check for Siemens Multitom RAX CT scanners with software version VF10. The model number 10860700 is associated with this recall.

  • Brand
    Siemens
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version on your Siemens Multitom RAX scanner. If it is VF10, it is affected.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA terminated this recall 2.6 years after the firm started it.

1. Firm started the recall: Apr 20, 202112. FDA classified it Class II: Jun 2, 202123. Published in FDA’s Enforcement Report: Jun 9, 202134. FDA terminated the recall: Nov 8, 2023420212023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1685-2021 (April 20, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1685-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA