The FDA announced on April 20, 2021, that Siemens Medical Solutions USA, Inc. is recalling Multitom RAX units with software version VF10 due to potential radiation exposure errors. The agency mandated the recall of 26 devices distributed nationwide.
How to tell if you have it
- Brand: Siemens
- Product: Multitom RAX
- Software version: VF10
- Model number: 10860700
The hazard
Siemens stated that the system may allow image acquisition when more than 20% of the beam misses the detector, potentially causing overshoot. In very rare cases, the calculated dose value may exceed the limit of 2 Gy due to incorrect source-to-image distance calculations, causing the buzzer to activate erroneously.
Context
GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Recent similar recalls include the Siemens ARTIS icono Ceiling Monitor Recall (June 8, 2026), the Siemens ARTIS icono floor recall (May 20, 2026), and the Siemens ARTIS Pheno VE30A/VE40A recall (March 12, 2026). Other recent recalls of similar products include Philips Ingenuity CT Scanners (August 7, 2026), Philips Brilliance CT 64 Medical Scanner Recall (August 7, 2026), and Philips IQon Spectral CT Software 4.7.7.x Recall (August 7, 2026).