Use caution Medical device recall · FDA · Published April 26, 2021

Xstrahl Concerto Software Recall: Risk of Mis-treatment with 2 Opposing Beams

Xstrahl recalls Concerto software versions V2.0-V2.2 for Xstrahl 100-300 systems. Editing treatment plans with 2 opposing beams before approval can cause mis-treatment errors. Affected users should contact Xstrahl for guidance.

🇺🇸 FDA recall #Z-0174-2022Medical deviceXstrahl ›
Take action

Contact the company

Recalling firmXstrahl LimitedUnit 2, Maybrook Industrial Estate Maybro; Maybrook Road · Walsall, United Kingdom
✎ GovInfoNow plain-language explainer

Xstrahl is recalling specific versions of its Concerto User Interface Software for certain radiation therapy systems. If users edit a treatment plan with two opposing beams before approval, the software may not update Beam 2 correctly, leading to potential mis-treatment during radiation therapy. Read more at govinfonow.com/recall/62727-xstrahl-concerto-software-recall-risk-of-mis-treatment-with-2.

Source Summary by GovInfoNow · govinfonow.com/recall/62727-xstrahl-concerto-software-recall-risk-of-mis-treatment-with-2 · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0174-2022

Cite this page

“Xstrahl Concerto Software Recall: Risk of Mis-treatment with 2 Opposing Beams.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62727-xstrahl-concerto-software-recall-risk-of-mis-treatment-with-2. Based on FDA recall #Z-0174-2022 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0174-2022%22&limit=1).

The FDA announced on April 26, 2021, that Xstrahl Limited is recalling Concerto User Interface Software provided with certain Xstrahl brachytherapy and superficial x-ray therapy systems. The software may fail to update beam parameters correctly, leading to error messages during treatment and possible mis-treatment.

How to tell if you have it

  • Xstrahl Concerto User Interface Software V2.0, V2.1, and V2.2
  • Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy System
  • Xstrahl 150 Electronic Brachytherapy - 150kV Superficial X-Ray Therapy System
  • Xstrahl 200 - 225kV Superficial / Orthovoltage X-Ray Therapy System
  • Xstrahl 300 - 300kV Orthovoltage X-Ray Therapy System
  • Xstrahl X 80 Photoelectric Therapy System

The hazard

If a saved treatment plan with two opposing beams is edited prior to approval, Beam 2 is not updated with the changed parameters upon selecting save. This results in error messages during the treatment and possible mis-treatment.

Context

GovInfoNow's records show Xstrahl Limited has issued two other recalls since August 27, 2021: a recall of the Xstrahl 150 Treatment Applicator on that date due to a base plate detachment hazard, and a recall of the Xstrahl X80 RADiant Therapy System on April 27, 2022, for compatibility issues with applicators. Other recent recalls of similar products include Elekta MOSAIQ Oncology System Software Recall (May 21, 2026) and Elekta Leksell GammaPlan software recalls (December 12, 2025).

How to tell if you have this one

Check if your Xstrahl system (100, 150, 200-225kV, 300kV, or X80) has Concerto software versions V2.0, V2.1, or V2.2. The software is included with these systems for radiation therapy.

  • Brand
    Xstrahl
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Identify your Xstrahl system model and the Concerto software version installed.

  2. Contact Xstrahl

    Reach out to Xstrahl for guidance on whether your software version is affected and what steps to take.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 9 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: named in the noticeORNevada: named in the noticeNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: named in the noticePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: named in the noticeKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: named in the noticeAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: named in the noticeTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Arizona, California, Florida, Kentucky, Nevada, Oregon, Pennsylvania, Tennessee, Texas.

Recall timeline

FDA terminated this recall 2.8 years after the firm started it.

1. Firm started the recall: Apr 26, 202112. FDA classified it Class II: Oct 26, 202123. Published in FDA’s Enforcement Report: Nov 3, 202134. FDA terminated the recall: Jan 26, 2024420212024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0174-2022 (April 26, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0174-2022 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA