Skip to content
Lower risk Medical device recall · FDA · Published May 10, 2021

BioFire BCID2 Panel Recall: False Negative Blood Culture Results Risk

BioFire recalls BCID2 Panel (REF RFIT-ASY-0147) due to potential false negatives in blood culture tests. Affected units have pouches with wrong packaging material. Contact for remedy.

🇺🇸 FDA recall #Z-1903-2021Medical deviceBioFire ›
Take action

Contact the company

Recalling firmBioFire Diagnostics, LLC515 S Colorow Dr · Salt Lake City, UT 84108-1248
✎ GovInfoNow plain-language explainer

BioFire is recalling its BCID2 Blood Culture Identification Panel because some units have pouches made with the wrong packaging material, which can cause false negative results in blood culture tests. This means the test might incorrectly show no infection when one exists. Read more at govinfonow.com/recall/62585-biofire-bcid2-panel-recall-false-negative-blood-culture-results-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/62585-biofire-bcid2-panel-recall-false-negative-blood-culture-results-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1903-2021

Cite this page

“BioFire BCID2 Panel Recall: False Negative Blood Culture Results Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62585-biofire-bcid2-panel-recall-false-negative-blood-culture-results-risk. Based on FDA recall #Z-1903-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1903-2021%22&limit=1).

The FDA announced on May 10, 2021, that BioFire Diagnostics, LLC voluntarily recalled the BioFire BCID2 Panel due to elevated false negative results. The recall affects 870 kits distributed nationwide in the United States.

How to tell if you have it

  • BioFire BCID2 Panel, REF RFIT-ASY-0147
  • UPC 00815381020338
  • Pouch Lot# 11YA20
  • Kit Lot# 2039020

The hazard

The recall was issued because pouches contained OPP from the FilmArray Blood Culture Identification Panel (BCID) instead of the correct BCID2 Panel, leading to elevated rates of false negative results. The FDA classified this as a Class II recall.

Context

GovInfoNow's records show BioFire Diagnostics, LLC has had 32 other recalls on record since March 14, 2018. Other recent recalls include the BioFire Respiratory Panel 2.1 (June 24, 2026), the BioFire Joint Infection Panel (April 22, 2026), and the BioFire Spotfire Respiratory Panel (April 1, 2026). Other recent recalls of similar products include the Biofire Defense FilmArray NGDS Warrior Panel (July 16, 2026), the Biomerieux VITEK 2 AST-N447 Kit (June 4, 2026), and the Biomerieux VITEK 2 AST-N461 Test Kit (July 24, 2026).

How to tell if you have this one

Check for the product code REF RFIT-ASY-0147 on the panel. Look for pouch lot number 11YA20 and kit lot number 2039020. The unique device identifier (UDI) is (01)00815381020338.

  • Brand
    BioFire
  • Product
    BioFire BCID2 Panel, REF RFIT-ASY-0147, For FilmArray Systems, CE, Blood Culture Identification Panel in vitro
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Panel Details

    Verify the product code REF RFIT-ASY-0147, pouch lot number 11YA20, and kit lot number 2039020 on the panel.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US nationwide distribution” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA terminated this recall 19 months after the firm started it.

1. Firm started the recall: May 10, 202112. FDA classified it Class II: Jun 21, 202123. Published in FDA’s Enforcement Report: Jun 30, 202134. FDA terminated the recall: Dec 15, 2022420212022
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1903-2021 (May 10, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1903-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA