Skip to content
Lower risk Medical device recall · FDA · Published June 10, 2021

GE OEC Medical X-Ray System Recall: Battery Depletion Risk

GE OEC recalls mobile fluoroscopy X-ray systems due to potential coin cell battery depletion causing system inoperability. Affected units may require replacement.

🇺🇸 FDA recall #Z-2160-2021Medical deviceGE OEC ›
Take action

Contact the company

Recalling firmGE OEC Medical Systems, Inc384 N Wright Brothers Dr · Salt Lake City, UT 84116-2862
✎ GovInfoNow plain-language explainer

GE OEC is recalling mobile fluoroscopy X:ray systems because the coin cell battery that monitors X-ray tube temperature may deplete early, causing the system to stop working. This could lead to interrupted medical procedures if the system becomes inoperable during critical treatments. Read more at govinfonow.com/recall/62354-ge-oec-medical-x-ray-system-recall-battery-depletion-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/62354-ge-oec-medical-x-ray-system-recall-battery-depletion-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2160-2021

Cite this page

“GE OEC Medical X-Ray System Recall: Battery Depletion Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62354-ge-oec-medical-x-ray-system-recall-battery-depletion-risk. Based on FDA recall #Z-2160-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2160-2021%22&limit=1).

GE OEC Medical Systems, Inc. recalled its OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile due to a potential early depletion of the coin cell battery that monitors X-ray tube temperature, which may cause the system to become inoperable, according to the FDA on June 10, 2021.

How to tell if you have it

  • GE OEC
  • OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile
  • All OEC 3D Systems

The hazard

The recall addresses the potential that the coin cell battery used to monitor X-ray tube temperature may deplete early, causing the system to become inoperable.

Context

GovInfoNow's records show GE OEC Medical Systems, Inc. has two other recalls on record since May 15, 2023. On November 21, 2024, the company recalled the GE OEC OEC 9800 X-Ray System due to an oil leak risk. Other recent recalls of similar products include the GE Optima XR220amx Mobile X-Ray System Recall (June 5, 2026) and the GE Medical Systems Allia IGS 3 Pulse X-Ray System Recall (April 20, 2026).

How to tell if you have this one

This recall applies to GE OEC's mobile fluoroscopy X-ray systems used for adult and pediatric imaging during procedures. The product is designed for diagnostic, interventional, and surgical use.

  • Brand
    GE OEC
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore, Australia, Thailand, United Kingdom, Italy, Korea, Russia, CHINA, Romania, Nigeria, Czechia, Argentina, Taiwan, Bangladeh, Malaysia, Bosnia and Herzegovina, Poland, Belgium, Indonesia, Dominican Republic, Switzerland, Hungary, New Zealand, Vietnam, Mexico, India, Norway, Philippines, Ecuador, Lithuania, HONDURAS, Uruguay, Sweden, Croatia, Jamaica, Kuwait, Ethiopia, Saudi Arabia, Finland, Bahamas, Trinidad and Tobago, Pakistan, Slovakia, Costa Rica, Israel, Panama, Greece, South Africa and DENMARK.
Recall numberProduct, codes & lotsQuantity
Z-2160-2021OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures; All OEC 3D Systems3 units
Z-2161-2021OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.; All OEC Elite systems manufactured on or after January 20193570 units
Based on the official FDA remedy

What to do — step by step

  1. Check System Status

    Verify if the system has a coin cell battery that monitors X-ray tube temperature.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 36 states.

Alaska: nationwideAKMaine: named in the noticeMEVermont: named in the noticeVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: named in the noticeIDMontana: named in the noticeMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: named in the noticeNVWyoming: named in the noticeWYSouth Dakota: named in the noticeSDIowa: named in the noticeIAIndiana: nationwideINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: nationwideNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: named in the noticeWVVirginia: nationwideVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: named in the noticeAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: named in the noticeDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: nationwideHITexas: nationwideTXFlorida: named in the noticeFL

FDA distribution: “Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Arizona, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, Nevada, New Jersey, North Carolina, Ohio, Oklahoma, Pennsylvania, South Carolina, South Dakota, Tennessee, Utah, Vermont, Washington, West Virginia, Wisconsin, Wyoming.

Recall timeline

FDA terminated this recall 2.9 years after the firm started it.

1. Firm started the recall: Jun 10, 202112. FDA classified it Class II: Jul 29, 202123. Published in FDA’s Enforcement Report: Aug 4, 202134. FDA terminated the recall: May 9, 2024420212024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2160-2021 (June 10, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2160-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA