Skip to content
Lower risk Medical device recall · FDA · Published June 30, 2021

GE Healthcare Revolution CT Systems: Incorrect Dose Display Risk

GE Healthcare recalls Revolution CT, Revolution CT ES, Revolution Apex, and Revolution CT with Apex edition CT scanners due to potential overdisplay of radiation dose during interventional exams with SmartStep.

🇺🇸 FDA recall #Z-0242-2022Medical deviceGE Healthcare ›
Take action

Contact the company

Recalling firmGE Healthcare, LLC3000 N Grandview Blvd · Waukesha, WI 53188-1615
✎ GovInfoNow plain-language explainer

GE Healthcare is recalling certain CT scanners because they may incorrectly show a much higher radiation dose than actually delivered during specific interventional exams. This could lead to unnecessary radiation exposure for patients if the dose is misinterpreted by medical staff. Read more at govinfonow.com/recall/62270-ge-healthcare-revolution-ct-systems-incorrect-dose-display-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/62270-ge-healthcare-revolution-ct-systems-incorrect-dose-display-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0242-2022

Cite this page

“GE Healthcare Revolution CT Systems: Incorrect Dose Display Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62270-ge-healthcare-revolution-ct-systems-incorrect-dose-display-risk. Based on FDA recall #Z-0242-2022 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0242-2022%22&limit=1).

The FDA announced on June 30, 2021, that GE Healthcare is recalling Computed Tomography Systems due to a hazard where accumulated radiation dose may be incorrectly displayed. The agency stated the error occurs during interventional exams if the table height is adjusted after the exam starts.

How to tell if you have it

  • GE Healthcare Revolution CT
  • GE Healthcare Revolution CT ES
  • GE Healthcare Revolution Apex
  • GE Healthcare Revolution CT with Apex edition

The hazard

The accumulated dose is incorrectly displayed in certain situations, showing a value up to 100 times higher than the actual dose delivered. This occurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts.

Context

GovInfoNow's records show GE Healthcare LLC has 97 other recalls on record since October 4, 2017. Other recent recalls of similar products include GE Healthcare SIGNA Premier MRI Scanners (February 27, 2026), GE Centricity Universal Viewer 6.0/7.0/8.0 (July 15, 2022), and another SIGNA Premier MRI Scanner recall (June 24, 2022). Other recent recalls of similar products include Philips Ingenuity CT Scanners (August 7, 2026), Philips Brilliance CT 64 Medical Scanner (August 7, 2026), and Philips Brilliance CT Scanners (August 7, 2026).

How to tell if you have this one

Check if your CT scanner model is Revolution CT, Revolution CT ES, Revolution Apex, or Revolution CT with Apex edition. These systems were sold with SmartStep option for interventional exams.

  • Brand
    GE Healthcare
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Scanner Model

    Identify your CT scanner model to confirm if it is one of the recalled Revolution CT, Revolution CT ES, Revolution Apex, or Revolution CT with Apex edition models.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide and US Nationwide Distribution” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 5 months after the firm started it.

1. Firm started the recall: Jun 30, 202112. FDA classified it Class II: Nov 24, 202123. Published in FDA’s Enforcement Report: Dec 1, 202132021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0242-2022 (June 30, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0242-2022 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA