The Food and Drug Administration announced on July 5, 2021, that Zeltiq Aesthetics, Inc. is recalling the CoolSculpting Elite System due to an incorrect error messaging system. The agency said the defect could lead users to re-treat an area within 24 hours or continue treating without knowing a thermal event occurred.
How to tell if you have it
- Brand: Zeltiq Aesthetics
- Product: CoolSculpting Elite System
- Software releases: SW release 2.0, SW release 3.0, SW release 3.0.1, SW release 3.0.2
- Model Numbers: CS-S3-002-D-00 and CS-S3-002-D-00_R (System); CS-A3-15S-00 (Surface 150 Applicator)
- UDI: 00816417021619 (System); 00816417021718 (Applicator)
- Units affected: 911 systems (updated to 1105 on Aug. 23, 2021)
The hazard
According to the FDA, an incorrect error messaging system could potentially lead to reporting a thermal event error causing a user to re-treat the affected anatomic area within 24 hours. It could also cause a user not to report a thermal event or any other error codes, leading them to continue treating without being aware that a thermal event has occurred.
Context
GovInfoNow's records show Zeltiq Aesthetics, Inc. has 1 other recall on record since 2022-06-09. On 2022-06-09, the company recalled CoolMax ZELTIQ Vacuum Applicators due to a risk of Paradoxical Hyperplasia. Other recent recalls of similar products include Elekta MOSAIQ Oncology System Recall (2025-12-10), Elekta MOSAIQ Oncology System Software Recall (2026-05-21), and GE Medical Systems Portrait Compute V1.2 Software Recall (2026-07-20).