The FDA announced on July 20, 2021, that Welch Allyn Inc is recalling Braun ThermoScan PRO 6000 Ear Thermometers due to overheating risks. The agency said the devices may cause burns if used before cleaning fluid dries completely.
How to tell if you have it
- Brand: Welch Allyn
- Product: Braun ThermoScan PRO 6000 Ear Thermometer
- Model: 6000
- UDI: 00732094309027 DDDYYKXXXXX (DDD = 3 digit Julian day, YY = last 2 digits of Julian year, K = Keytronic, XXXXX = 5 digit serialized number)
- Units affected: 1,036,740
The hazard
According to the FDA, if the device is exposed to fluid ingress and used before the cleaning fluid has completely dried, there is a risk of the device overheating. This overheating potentially causes a burn to the user or the ear canal of the patient.
Context
GovInfoNow's records show Welch Allyn Inc had one other recall on record since April 22, 2021, involving the Welch Allyn ProBP 3400 Blood Pressure Device due to electrical shock risks from fluid ingress. Other recent recalls of similar products include BD Pyxis Anesthesia Station ES REF 323 Medication Cabinet Recall (April 30, 2026), BD Pyxis MedStation ES Tower 352 Medication Cabinet Recall (April 30, 2026), and BD Pyxis Anesthesia Station ES REF 327 Medication Cabinet Recall (April 30, 2026).