Abbott Alinity ci-series System Control Module Software Version 3.2.3 and Earlier Recalled

Abbott recalls Alinity ci-series System Control Module units with software versions 3.2.3 and earlier due to potential performance issues including incorrect reagent status displays, damaged connectors, and sample processing errors. Affected units have specific serial numbers and model numbers.

🇺🇸 FDA recall #Z-2472-2021Medical deviceAbbott ›
Take action

Contact the company

Recalling firmAbbott Laboratories1915 Hurd Dr · Irving, TX 75038
✎ GovInfoNow plain-language explainer

What happened, in plain English

Abbott is recalling Alinity ci-series System Control Modules with software versions 3.2.3 and earlier because they may display expired reagents as 'OK', cause damaged connectors during maintenance, or lead to incorrect test results. This could affect lab results and equipment safety. Read more at govinfonow.com/recall/62155-abbott-alinity-ci-series-system-control-module-software-version-3.

Source Summary by GovInfoNow · govinfonow.com/recall/62155-abbott-alinity-ci-series-system-control-module-software-version-3 · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2472-2021

Cite this page

“Abbott Alinity ci-series System Control Module Software Version 3.2.3 and Earlier Recalled.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/62155-abbott-alinity-ci-series-system-control-module-software-version-3. Based on FDA recall #Z-2472-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2472-2021%22&limit=1).

Why it’s dangerous

The software issues can cause the system to incorrectly show expired reagents as 'OK', leading to expired reagents being used. Performing certain maintenance procedures may damage connectors, and sample processing errors could result in incorrect test results without proper warnings.

Who’s affected

Healthcare labs and clinical testing facilities that use the Alinity ci-series System Control Module with software versions 3.2.3 and earlier are most at risk. Individuals with the affected units should check their serial numbers.

How to tell if you have this one

Check if your Alinity ci-series System Control Module has software version 3.2.3 or earlier. Affected units have model number LN 03R70-01 and serial numbers starting with SCM (e.g., SCM01001 to SCM04259).

  • Brand
    Abbott
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your software version by checking the device settings or documentation. If it's 3.2.3 or earlier, it's affected.

  2. Contact Abbott

    Reach out to Abbott Laboratories for instructions on how to update the software or return the device if needed.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Worldwide distribution. US nationwide including Puerto Rico; Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Canada, Chile,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 58 days after the firm started it.

1. Firm started the recall: Jul 26, 202112. FDA classified it Class II: Sep 14, 202123. Published in FDA’s Enforcement Report: Sep 22, 202132021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Abbott recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2472-2021 (July 26, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2472-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA