The FDA announced on 2021-09-29 that Covidien LLC voluntarily recalled the Emprint Ablation Catheter with Thermosphere Technology due to a temperature monitoring circuitry error. This defect could cause the applicator tip to overheat, potentially resulting in heat damage and allergic or toxic reactions.
How to tell if you have it
- Brand: Covidien
- Product: Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation
- CFN: CA108L1
- Lot code: 516840
The hazard
The firm identified an error in the device temperature monitoring circuitry designed to interrupt energy delivery during a primary cooling system failure. Use of an affected device with such a failure could lead to applicator tip overheating and associated heat damage, which could result in an allergic/toxic reaction, unintended tissue ablation, tissue burn, and/or a delay of treatment while an alternate device is located.