Skip to content
Lower risk Medical device recall · FDA · Published October 13, 2021

HemosIL ReadiPlasTin (20mL) recall: Risk of inaccurate test results

Instrumentation Laboratory recalls HemosIL ReadiPlasTin (20mL) lot N0705526 and N0806224 due to potential inaccurate test results. Affected users should check their product labels.

🇺🇸 FDA recall #Z-0276-2022Medical deviceInstrumentation Laboratory ›
Take action

Contact the company

Recalling firmInstrumentation Laboratory180 Hartwell Rd · Bedford, MA 01730-2443
✎ GovInfoNow plain-language explainer

Instrumentation Laboratory is recalling HemosIL ReadiPlasTin (20 mL) reagent lots N0705526 and N0806224 because customer reports show increased imprecision and potential for inaccurate test results. This could lead to incorrect medical conclusions if quality controls aren't properly followed. Read more at govinfonow.com/recall/61689-hemosil-readiplastin-20ml-recall-risk-of-inaccurate-test-results.

Source Summary by GovInfoNow · govinfonow.com/recall/61689-hemosil-readiplastin-20ml-recall-risk-of-inaccurate-test-results · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0276-2022

Cite this page

“HemosIL ReadiPlasTin (20mL) recall: Risk of inaccurate test results.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/61689-hemosil-readiplastin-20ml-recall-risk-of-inaccurate-test-results. Based on FDA recall #Z-0276-2022 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0276-2022%22&limit=1).

The FDA announced on 2021-10-13 that Instrumentation Laboratory is recalling HemosIL ReadiPlasTin (20 mL) due to performance issues that may cause erroneous test results. The agency issued recall number Z-0276-2022 for the in vitro diagnostic thromboplastin reagent.

How to tell if you have it

  • Brand: Instrumentation Laboratory
  • Product: HemosIL ReadiPlasTin (20 mL)
  • Part number: 0020301400
  • Lot codes: N0705526, N0806224
  • UDI: 08426950632887

The hazard

The firm has received customer reports of performance issues with the affected lots, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Context

GovInfoNow's records show Instrumentation Laboratory has 13 other recalls on record since 2021-08-25. Recent brand recalls include the GEM Premier 5000 PAK lysing bag solution risk (2026-05-21), the GEM Premier 5000 blood analyzer wrong patient ID assignment risk (2026-05-08), and GEM PAKs causing repeated warm-up errors (2026-01-20). Other recent recalls of similar products include the Siemens Dimension Tacrolimus Flex Reagent Cartridge Recall (2026-05-07), the Siemens Healthcare Syva EMIT 2000 Quinidine Assay Recall (2026-07-13), and the Siemens Dimension Creatinine Flex Reagent Cartridge Recall (2026-03-18).

How to tell if you have this one

Check for lot numbers N0705526 or N0806224 on the product label. The product is an in vitro diagnostic thromboplastin reagent sold by Instrumentation Laboratory.

  • Brand
    Instrumentation Laboratory
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Product Label

    Look for lot numbers N0705526 or N0806224 on the HemosIL ReadiPlasTin (20 mL) reagent label.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 12 states.

Alaska: nationwideAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: named in the noticeILWisconsin: nationwideWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: nationwideINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: nationwideMDDelaware: nationwideDEArizona: named in the noticeAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: nationwideDCOklahoma: named in the noticeOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: nationwideTXFlorida: nationwideFL

FDA distribution: “Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. ***Updated 3/3/22*** Expansion…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Arizona, California, Illinois, Iowa, Kansas, Maine, Michigan, New York, North Carolina, Oklahoma, South Carolina, Virginia.

Recall timeline

FDA published this recall 49 days after the firm started it.

1. Firm started the recall: Oct 13, 202112. FDA classified it Class II: Nov 24, 202123. Published in FDA’s Enforcement Report: Dec 1, 202132021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0276-2022 (October 13, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0276-2022 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA