Skip to content
Lower risk Medical device recall · FDA · Published October 15, 2021

Freelite Human Lambda Free Kit Recall: Calibration Issues May Cause Inaccurate Results

The Binding Site Group recalls Freelite Human Lambda Free Kit (REF LK018.S) due to calibration issues that may cause inaccurate test results. Affected units include specific lot numbers and model numbers.

🇺🇸 FDA recall #Z-0274-2022Medical deviceThe Binding Site Group ›
Take action

Contact the company

Recalling firmThe Binding Site Group, Ltd.8 Calthorpe Road · Birmingham, United Kingdom
✎ GovInfoNow plain-language explainer

The Binding Site Group is recalling Freelite Human Lambda Free Kit (for use on SPAPLUS) because calibration problems could lead to inaccurate test results. This affects specific model numbers and lot numbers, and users should check their product details. Read more at govinfonow.com/recall/61673-freelite-human-lambda-free-kit-recall-calibration-issues-may-cause.

Source Summary by GovInfoNow · govinfonow.com/recall/61673-freelite-human-lambda-free-kit-recall-calibration-issues-may-cause · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0274-2022

Cite this page

“Freelite Human Lambda Free Kit Recall: Calibration Issues May Cause Inaccurate Results.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/61673-freelite-human-lambda-free-kit-recall-calibration-issues-may-cause. Based on FDA recall #Z-0274-2022 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0274-2022%22&limit=1).

The Binding Site Group, Ltd. recalled Freelite Human Lambda Free Kits on Oct. 15, 2021, due to calibration curve errors that may produce higher activity than expected, the FDA said in a Class II recall.

How to tell if you have it

  • Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S, Lot Number: 472776
  • Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S, Lot Number: 472664

The hazard

The recall cites calibration curves producing higher activity than routinely expected and a potential decrease in separation between calibrator points 5 and 6, which can be exacerbated when calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.

Context

GovInfoNow's records show The Binding Site Group, Ltd. has had 15 other recalls on record since Aug. 23, 2017. Other recent recalls of similar products include EXENT Analyser Software Version 1.0.20 (April 16, 2026) and Freelite Human Kappa Free Kit (May 20, 2021).

How to tell if you have this one

Check for model numbers LK018.S or LK018.10S and lot numbers 472776 or 472664 on the product packaging or label.

  • Brand
    The Binding Site Group
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Product Details

    Identify your model number and lot number to confirm if you have an affected unit.

Where it went and when

Where it was distributed

Distributed in 13 states, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: yesILWisconsin: not listedWIMichigan: yesMINew York: yesNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: yesSDIowa: not listedIAIndiana: yesINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: yesCAUtah: not listedUTColorado: yesCONebraska: not listedNEMissouri: not listedMOKentucky: yesKYWest Virginia: not listedWVVirginia: yesVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: yesOKLouisiana: yesLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: yesTXFlorida: yesFL

FDA distribution: “U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: Canada” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: California, Colorado, Florida, Illinois, Indiana, Kentucky, Louisiana, Michigan, New York, Oklahoma, South Dakota, Texas, Virginia.

Recall timeline

FDA terminated this recall 3.3 years after the firm started it.

1. Firm started the recall: Oct 15, 202112. FDA classified it Class II: Nov 19, 202123. Published in FDA’s Enforcement Report: Dec 1, 202134. FDA terminated the recall: Feb 4, 2025420212025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using the kit immediately?

The recall does not require immediate stop use; check your product details to see if you are affected.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0274-2022 (October 15, 2021) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0274-2022 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA