The FDA announced on 2021-11-09 that Medtronic Navigation, Inc. is recalling Stealth Autoguide Trackers due to potential weld separation. This defect may cause navigation inaccuracy and tissue injury during neurosurgery.
How to tell if you have it
- Brand: Medtronic
- Product: Stealth Autoguide Tracker, Model: 28248
- Part of Stealth Autoguide Basic Instrument Kit, Model: 9736188
- All lot numbers
The hazard
A tracker that holds and guides neurosurgical instruments may suffer weld failure where it connects to the instrument tube. If separation occurs, navigation inaccuracy may result in prolonged procedure and tissue injury.
Context
GovInfoNow's records show Medtronic Navigation, Inc. has 25 other recalls on record since 2017-06-07. Recent issues include 2024-12-04 surgical pins that may not fit patient reference frames and 2024-08-22 StealthStation S8 App software anomalies causing flipped images. Other recent recalls of similar products include Medtronic Bio-Medicus Adult Cannula Kit Recall (2026-05-25), Boston Scientific ENROUTE Transcarotid Neuroprotection System Recall (2026-07-09), and Alphatec LIF NAV Osteotome 10mm (2026-01-02).