The FDA announced on 2021-11-11 that Medtronic Navigation, Inc. is recalling Stealthstation Systems with specific cranial software versions due to a hazard of incorrect needle position display. This error may result in prolonged procedures, additional surgeries, or tissue injury.
How to tell if you have it
- Stealthstation System w/ Stealthstation Cranial Software 3.1.1, 3.1.2, 3.1. or SynergyCranial Software
- Synergy Cranial, Model: 9733763
- StealthStation Cranial, Model: 9735585; 9733763, Version: 2.2.8
- 9735585, Versions: 3.1.1, 3.1.2, 3.1.3
The hazard
According to the FDA, the cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen. This may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, or tissue injury.
Context
GovInfoNow's records show Medtronic Navigation, Inc. has 25 other recalls on record since 2017-06-07. Recent brand recalls include issues with PIN surgical pins not fitting patient reference frames (2024-12-04), flipped images in StealthStation S8 App versions (2024-08-22), and missing digits during neurosurgery (2024-03-26). Other recent recalls of similar products include Howmedica Stryker CranialMask Tracker software errors (2025-11-12), Medtronic Vanta CP App software risks (2025-12-03), and Hitachi Proton Beam Therapy System positioning anomalies (2026-02-06).