The FDA announced on November 15, 2021, that Datascope Corp. is recalling Cardiosave Rescue IntraAortic Balloon Pumps due to the possibility of fluid ingress. Fluid entering the device may short electronic components and cause system shutdown, leading to unsafe hemodynamic instability.
How to tell if you have it
- Brand: Datascope
- Product: Cardiosave Rescue IntraAortic Balloon Pump
- Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83, 0998-00-0800-85, 0998-UC-0800-85
- Serial Numbers: All Serial Numbers
- Part Number: D998
- Distribution: US Nationwide
The hazard
The recall addresses the possibility of fluid entering the device, which may short various electronic components and lead to system shutdown. This failure could result in unsafe hemodynamic instability for patients using the pump.
Context
GovInfoNow's records show Datascope Corp. has 29 other recalls on record since June 17, 2019. Other recent recalls of similar products include Boston Scientific Intera 3000 Hepatic Artery Pump Recall (May 11, 2026), Fresenius Kabi Ivenix Infusion Pump Recall (November 3, 2025), and ICU Medical Plum Solo IV Pump Recall (April 29, 2026).