Philips North America Llc recalled Philips Allura Xper FD series medical imaging systems due to the risk of monitor components falling and causing injury, according to the FDA. The agency announced the recall on November 19, 2021.
How to tell if you have it
- Brand: Philips
- Product: Philips Allura Xper FD series with Software Version Number: 2.1.x-
- Model Numbers: 722005, 722006, 722008, 722010, 722012, 722013, 722025, 722026, 722027, 722028
- UPCs: 00884838054189, 00884838054196, 00884838054202, 00884838059030, 00884838059054, 00884838059191
- Distribution: Worldwide; US states of DC, GA, MA, TX, VA; countries of Dominican Republic, India, Japan, Poland, Spain, Taiwan
The hazard
Set screws securing the gliding rods of Extra Monitors can come loose, causing the monitors to disengage from the Mounting Ceiling Suspension structure and fall. This may lead to harm to patients, users, or service engineers that requires medical intervention.
Context
GovInfoNow's records show Philips North America Llc has 129 other recalls on record since July 12, 2017. Other recent recalls of similar products include Evolution upgrade 1.5T and Evolution upgrade 3.0T (August 21, 2026), Spectral CT 7500 Scanner (August 18, 2026), and Ingenuity CT Scanners (August 7, 2026).