The FDA announced on Dec. 22, 2021, that Covidien, LP voluntarily recalled the SonarMed AirWave Monitor Kit (Item Code AW-M0001) due to alarm and sensor detection failures. The agency issued recall number Z-0514-2022 for the Class II device.
How to tell if you have it
- Brand: Covidien
- Product: SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord)
- Item Code: AW-M0001
- GTIN/UPN: 00851334007001
- Serial Numbers: AW0501, AW0519, AW0537, AW0556, AW0502, AW0520, AW0540, AW0557, AW0503, AW0521, AW0541, AW0558, AW0504, AW0522, AW0542, AW0559, AW0505, AW0524, AW0543, AW0560, AW0506, AW0525, AW0544, AW0561, AW0507, AW0526, AW0545, AW0562, AW0508, AW0527, AW0546, AW0563, AW0509, AW0528, AW0548, AW0564, AW0510, AW0529, AW0549, AW0565, AW0511, AW0530, AW0550, AW0566, AW0512, AW0531, AW0551, AW0567, AW0513, AW0532, AW0552, AW0568, AW0514, AW0533, AW0553, AW0569, AW0515, AW0536, AW0554, AW0570, AW0516, AW0538, AW0555, AW0571, AW0517, AW0539
- Quantity: 66 devices
- Distribution states: CA, CO, FL, IL, KY, MA, OH, TN, TX
The hazard
GovInfoNow's records show the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.
Context
GovInfoNow's records show Covidien, LP has 26 other recalls on record since 2017-07-17. Other recent recalls of similar products include Philips IntelliVue MX800 Patient Monitor Recall: No Alarms May Cause Missed Alerts (2025-10-31), Philips IntelliVue MX450 Patient Monitor Recall: No Alarms May Cause Missed Alerts (2025-10-31) and Philips IntelliVue MX550 Patient Monitor Recall: No Alarms May Cause Missed Alerts (2025-10-31).