The FDA announced on Jan. 20, 2022, that Ventana Medical Systems Inc. is recalling BenchMark ULTRA and DISCOVERY ULTRA Instruments due to a potential fluid leak that could cause an electrical short circuit, smoke, and fire. Owners should contact the company for remedies.
How to tell if you have it
- Brand: Ventana Medical Systems
- Products: BenchMark ULTRA and DISCOVERY ULTRA Instruments
- Serial Numbers: Less-Than-Or-Equal-To 322030
The hazard
The recall addresses a potential for fluid leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector. This malfunction could result in smoke and fire, according to the FDA.
Context
GovInfoNow's records show Ventana Medical Systems, Inc. has 7 other recalls on record since Dec. 19, 2017. Other recent recalls of similar products include Ventana navify Digital Pathology v2.5 Software Image Discrepancy Recall (April 20, 2026), Ventana anti-CD10 antibody recall (June 27, 2024), and Ventana PD-L1 (SP142) Assay Recall (Nov. 18, 2022).