Philips Respironics recalls V60 Plus Ventilators with part number 1138747 due to expired adhesive that may cause silent shutdowns. Affected units have specific serial numbers listed in the recall notice.
A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).
Philips Respironics is recalling a limited number of V60 Plus Ventilators (part number 1138747) because some units were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and damage capacitors, potentially causing the ventilator to stop providing ventilation without warning (silent shutdown).
The ventilator uses expired adhesive in some units, which can fail over time. If the adhesive fails, a bracket may become loose and damage capacitors inside the device. This could cause the ventilator to stop providing ventilation to the patient without triggering alarms (silent shutdown) or it might activate both visual and audible alarms.
Patients using the Philips Respironics V60 Plus Ventilator with part number 1138747 and specific serial numbers listed in the recall notice are most at risk. The hazard could cause the ventilator to stop working without alerting the user.
Check if your ventilator has part number 1138747 and serial numbers listed in the recall notice (MX0016240 through MX0018224). The affected units are limited to specific models and serial numbers.
Verify your ventilator's part number is 1138747 and serial number matches the list of affected units (MX0016240 through MX0018224).
If your ventilator is affected, contact Philips Respironics for instructions on how to address the issue.
Philips Respironics is recalling these ventilators because a limited number were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and damage capacitors, potentially causing the ventilator to stop providing ventilation without warning.
The ventilator may stop providing ventilation without triggering alarms (silent shutdown) or it might activate both visual and audible alarms. This could be dangerous for patients who rely on the device.
Check if your ventilator has part number 1138747 and a serial number in the list provided (MX0016240 through MX0018224).
We’ve drafted a message you can send Philips Respironics to request your refund or repair — edit it as you like.
Subject: Recall Z-0663-2022 — Philips Respironics Philips Respironics Hello, I own a Philips Respironics that is covered by recall Z-0663-2022 from Philips Respironics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.