Bolder Surgical, LLC voluntarily recalled the CoolSeal Generator (REF: CSL-200-50) on Jan. 25, 2022, according to the FDA. The device may deliver unintended energy to patients due to a manufacturing defect. Owners should contact the company for a replacement.
How to tell if you have it
- Brand: Bolder Surgical
- Product: CoolSeal Generator
- REF: CSL-200-50
- Serial Numbers: 00161-170
- UDI: 00850346007023
The hazard
A potential manufacturing or assembly defect could cause the generator to fail IEC 60601 standards. This defect may lead to the delivery of unintended energy to the patient under specific conditions.
Context
GovInfoNow's records show other recent recalls of similar medical devices include the Philips Allura Xper FD10F Medical System Recall (2025-12-15) and the Philips Allura Xper FD10C Medical System Recall (2025-12-15). Additional recent recalls include the GE Healthcare Carestation Anesthesia System Power Board Recall (2025-11-14).