B Braun Medical recalls Conductivity Sensors (VERSION 2) for Dialog and Dialog+ hemodialysis systems due to potential cracks causing leaks that may lead to improper ultrafiltration. Affected units include specific model numbers and lots.
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
B Braun Medical is recalling Conductivity Sensors (VERSION 2) used with Dialog and Dialog+ hemodialysis machines because cracks could cause leaks in the dialysis circuit. This might lead to either too much or too little ultrafiltration, which could affect treatment safety.
Cracks in the conductivity sensors during use could cause leaks in the dialysate circuit. This might lead to either excessive or insufficient ultrafiltration depending on whether the machine has a Dialysis Fluid filter.
Hemodialysis patients using Dialog or Dialog+ machines with the recalled Conductivity Sensors (VERSION 2) are most at risk. The affected units include specific model numbers and production lots.
Check for Conductivity Sensor (END) - VERSION 2 with Article Code 3456103A. Affected models include 710200L, 710200S, 710200U, 710200K, 710500L, and 710500K. Production lots 4/21 and 5/21 are also affected.
Inspect the conductivity sensor for visible cracks or damage before use.
Reach out to B Braun Medical for guidance on whether the sensor needs replacement.
No, the recall is for specific conductivity sensors, not the entire machine. Continue using the machine as normal but check the sensor for cracks.
Cracks could cause leaks in the dialysate circuit, potentially leading to either excessive or insufficient ultrafiltration depending on the machine's filter setup.
Look for Conductivity Sensor (END) - VERSION 2 with Article Code 3456103A on models 710200L, 710200S, 710200U, 710200K, 710500L, or 710500K from production lots 4/21 and 5/21.
We’ve drafted a message you can send B Braun Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0873-2022 — B Braun Medical B Braun Medical Hello, I own a B Braun Medical that is covered by recall Z-0873-2022 from B Braun Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.