The FDA announced on Feb. 2, 2022, that Philips North America LLC is recalling Philips Allura and Philips Azurion medical devices due to a software defect. The hazard involves the StentBoost Live R2.0 software potentially displaying X-ray images from previous runs instead of current ones.
How to tell if you have it
- Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035)
- Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228)
- StentBoost Live (Model Number 459801781661) Software Version Number: R2.0
- UPCs: 00884838054189, 00884838054196, 00884838054202, 00884838054219, 00884838054240, 00884838059030, 00884838059054, 00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210
- Global Distribution. US Nationwide.
- 338 units affected.
The hazard
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Context
GovInfoNow's records show Philips North America LLC has had 129 other recalls on record since July 12, 2017. Other recent recalls of similar products include Philips Azurion System Software R3.1 May Cause Imaging Loss (Dec. 12, 2025) and Philips Azurion Systems R3.0 Software Issues: Risk of Image Loss and System Failures (Dec. 8, 2025).