The FDA announced on Feb. 22, 2022, that Philips North America Llc is recalling its Hemodynamic Application, Model Number 722463, due to a hazard involving incorrect blood pressure settings for neonatal patients. The agency classified the recall as Class II.
How to tell if you have it
- Brand: Philips
- Product: Philips Hemodynamic Application
- Model Number: 722463
- Software Version Number: R1.2.2 and R1.2.3
- UDI: B-00884838084544
- System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 886 14 15 643 1194 220 1238 947 949 950 951 89 657 169 168 413 414 624 565 1060 393 512 278 818 362 123 802 395 576 932 291 834 539 605 161 970 686 761 229 230
The hazard
According to the FDA, when a Neonatal patient is admitted, the Non-Invasive Blood Pressure may not be set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, nerve damage or bone deformity for a neonate.
Context
GovInfoNow's records show Philips North America Llc has 129 other recalls on record since 2017-07-12. Other recent recalls include SmartPath to dStream recalls (2026-08-21) and Spectral CT 7500 Scanner recalls (2026-08-18). Similar product recalls in the sector include Medline Neonatal Breathing Circuits risks reported on 2026-07-13 and Hamilton Medical IntelliCuff Recall issues reported on 2026-05-22.