The FDA announced on March 7, 2022, that Siemens Medical Solutions USA, Inc. is recalling SIEMENS SOMATOM computed tomography x-ray systems due to a software error. The defect may cause scanning workflow interruptions and image artifacts.
How to tell if you have it
- Brand: Siemens
- Product: SIEMENS SOMATOM with software syngo CT VA30A_SP4
- Models: SOMATOM go.Now (11061610, 11061618), SOMATOM go.Up (11061620, 11061628), SOMATOM go.All (11061630, 11061638), SOMATOM go.Top (11061640, 11061648), SOMATOM go.Sim (11061660, 11061668), SOMATOM go.Open Pro (11061670, 11061678), SOMATOM X.cite (11330001)
- Serial Numbers: 106115 through 128102 and 130012 through 130065, 168108, 168163, 172508 through 172550
- UDIs: 04056869051291, 04056869051307, 04056869051314, 04056869051321, 04056869151564, 04056869151588, 04056869151571, 04056869151595, 04056869249230, 04056869239736, 04056869249247, 04056869239743, 04056869231044
- Units Affected: 547 units in the US
The hazard
According to the FDA, a software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications, and image artifacts. These errors may also occur during interventional workflows, potentially leading to patient issues such as rescan requirements, unexpected X-ray dose, additional contrast media, delay in diagnosis, or scan abort.
Context
GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Recent brand recalls include the Siemens ARTIS icono Ceiling Monitor Recall dated June 8, 2026, for loose screws, and the Siemens ARTIS Icono Biplane Recall dated May 20, 2026, for risk of movement loss. Other recent recalls of similar products include Philips Spectral CT 7500 RT Software Issues May Cause Image Artifacts (August 21, 2026) and Philips Ingenuity CT Scanners: Unintended Radiation Risk (August 7, 2026).