The FDA announced on April 7, 2022, that Bio-Rad Laboratories is recalling specific lots of the VARIANT nbs Sickle Cell Program Reorder Pack due to software errors. The hazard involves custom settings being overwritten by default parameters during updates.
How to tell if you have it
- Brand: Bio-Rad
- Product: VARIANT nbs Sickle Cell Program Reorder Pack, REF: 250-3000
- Component: Resin UPDATE CD-ROM, REF: 250-3020
- REF 250-3020 Lot Codes: AX90085, AY90085, AZ90085, BB90085, CB90085, DB90085
- REF 250-3000 Lot/Expiry Codes: 64442741/2/28/2023, 64444208/2/28/2023, 64444218/3/26/2023, 64444715/3/31/2023, 64452929/3/31/2023, 64452922/5/31/2023
- Distribution: US states AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR; Canada, Spain, France, Italy, India, Vietnam, Philippines
The hazard
According to the FDA, the software causes customized settings in the Setup/Test/Data Setup and Setup/Test/Pattern Setting subscreens to be overwritten with default settings when the Update Kit procedure is performed. This risk arises because customer custom pattern settings may differ from default settings, potentially resulting in different patterns being assigned to validated samples. The firm initiated this recall following complaints from four customers who experienced these issues.
Context
GovInfoNow's records show Bio-Rad Laboratories has 12 other recalls on record since April 2, 2018. Other recent recalls of similar products include the Bio-Rad Platelia Toxo IgM Kit Recall (May 30, 2025) and the Bio-Rad BioPlex 220 Recall (Feb. 28, 2023).