The FDA announced on May 24, 2022, that CareFusion 303, Inc. is recalling BD Pyxis MedBank Cabinet Software v2.1.2.9 for legacy cabinets with Matrix drawers due to a software incompatibility hazard. The recall affects 686 units distributed nationwide.
How to tell if you have it
- Brand: BD Pyxis
- Product: BD Pyxis MedBank Cabinet Software v2.1.2.9
- Affected Systems: Legacy MedBank Cabinets with Matrix Drawers, including CUBEX MedFlex, MedBank 200 Main, and MedBank 500 Main operating with Windows 7
- Units Affected: 686 cabinets
The hazard
The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning, causing Matrix drawers to stop opening in the MedBank Application, which leads to delays in access to medications or supplies needed in residential or procedural care facilities.
Context
GovInfoNow's records show CareFusion 303, Inc. has 64 other recalls on record since June 12, 2017. Other recent recalls of similar products include a fingerprint scanner overheating hazard for the BD Pyxis MedBank Tower 1145-00 (April 30, 2026) and a software update issue for the BD Pyxis ES Enterprise Server (December 3, 2025).