The FDA announced on May 31, 2022, that Spacelabs Healthcare is recalling the Xhibit Telemetry Receiver, Model 96280. The device has a software defect that may prevent technical alarms from escalating after approximately 25 days of continuous use.
How to tell if you have it
- Brand: Spacelabs Healthcare
- Product: Xhibit Telemetry Receiver
- Model: 96280
- Software version: 1.4.0
- UDI-DI: 10841522100246
- Units affected: 529
The hazard
According to the FDA, a software-related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarms for signal loss, all leads off, low battery, SpO2 sensor off, and signal interference do not escalate as specified. This may cause caregivers to miss initial technical alarms and remain unaware of continuing technical failures.
Context
GovInfoNow's records show Spacelabs Healthcare has issued 10 other recalls since January 17, 2018. Recent Spacelabs recalls include the Xhibit Central Station Model 96102 audio/video display failures (June 4, 2026), the Ultraview SL Command Module electrical leakage risk (March 26, 2026), and the Multi-parameter Command Module Model 91496 inaccurate cardiac output readings (February 27, 2026). Other recent recalls of similar products include the GE HealthCare ApexPro v6.0 eDelivery Software recall for potential ECG and SpO2 monitoring loss (April 24, 2026), the GE HealthCare CARESCAPE Telemetry Server ARK-2250L recall for potential ECG/SpO2 monitoring loss (April 24, 2026), and the Philips IntelliVue MX70 monitor may not alarm recall (October 31, 2025).