Alere San Diego, Inc. recalled The Cholestech LDX System cassettes on June 24, 2022, according to the FDA. The hazard involves potential inaccurate test results due to bilirubin interference. Owners should update package inserts with new information.
How to tell if you have it
- Brand: Alere
- Product: Cholestech LDX System cassettes
- Part Numbers: 10-991, 99021, 14-531, 10-990, 10-989, 10-987
- Lot Codes/UDI-DI: 00893038002654, 00899722002344, 04571226470831, 00893038002647, 00893038002630, 00893038002616
- Units affected: 436,990
The hazard
Samples containing bilirubin at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated) can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Context
GovInfoNow's records show Alere San Diego, Inc. has had four other recalls since November 14, 2017. Other recent recalls of similar products include Beckman Coulter LDL Cholesterol Reagent May Give False High Results (December 8, 2025), Instrumentation Laboratory GEM Premier 5000 PAK: Lysing Bag Solution Risk (May 21, 2026), and Accriva Diagnostics recalls directCHECK ACT-LR blood control with incorrect labeling (April 30, 2026).