The FDA announced on July 19, 2022, that Ortho-Clinical Diagnostics is recalling VITROS XT 7600 Integrated Systems with Version 3.7.2 and below due to potential shifts in test results. The hazard stems from system shutdowns affecting the electrical current supplied to the Digital Imaging Reflectometer LED.
How to tell if you have it
- Brand: Ortho-Clinical
- Product: VITROS XT 7600 Integrated System with Version 3.7.2 and below
- Product Code: 6844461
- UDI-DI: 10758750031610
- System start codes: 7600XXXX where XXXX is unique for each analyzer
The hazard
Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used with the Digital Imaging Reflectometer. This potential shift can result in either a positive or negative shift in results reported by the VITROS System.
Context
GovInfoNow's records show Ortho-Clinical Diagnostics, Inc. has 53 other recalls on record since July 10, 2017. Other recent recalls of similar products include Ortho-Clinical LAC Slides Recall (July 29, 2025), Ortho-Clinical VITROS Ca Slides Recall (July 9, 2025), and Ortho-Clinical VITROS OP Reagent Gen 15 May Cause Inaccurate Test Results (June 12, 2025). Other recent recalls of similar products include Vivoo pH Test Recall (Nov. 22, 2025), Vivoo Vaginal pH Test Recall (Nov. 22, 2025), and Vivoo Test Strips Recall (Nov. 22, 2025).