Use caution Medical device recall · FDA · Published September 21, 2022

Microvention WEB Detachment Controller Recall: Risk of Implant Misplacement

Microvention recalls WEB Detachment Controller (WDC-2) due to potential implant misplacement during aneurysm embolization. Affected units have specific serial numbers sold in China and Japan. Contact for remedy.

🇺🇸 FDA recall #Z-0170-2023Medical deviceMicrovention ›
Take action

Contact the company

Recalling firmMICROVENTION INC.35 Enterprise · Aliso Viejo, CA 92656
✎ GovInfoNow plain-language explainer

Microvention is recalling its WEB Detachment Controller (model WDC-2) because a manufacturing defect may cause the implant device to move and damage blood vessels during surgery. This could lead to serious complications if the implant is not properly detached. Read more at govinfonow.com/recall/60032-microvention-web-detachment-controller-recall-risk-of-implant-misplacement.

Source Summary by GovInfoNow · govinfonow.com/recall/60032-microvention-web-detachment-controller-recall-risk-of-implant-misplacement · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0170-2023

Cite this page

“Microvention WEB Detachment Controller Recall: Risk of Implant Misplacement.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/60032-microvention-web-detachment-controller-recall-risk-of-implant-misplacement. Based on FDA recall #Z-0170-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0170-2023%22&limit=1).

The FDA announced on 2022-09-21 that Microvention Inc. is recalling WEB Detachment Controllers due to a defect that may prevent proper detachment of implant devices. The agency classified the recall as Class II and ongoing.

How to tell if you have it

  • Brand: Microvention
  • Product: WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2
  • Component of WEB Aneurysm Embolization System
  • Units affected: 2018

The hazard

The detachment controller has an out-of-specification tunnel opening dimension that may prevent the implant device pusher from being fully inserted. This could result in the inability to properly detach the implant device, potentially causing delays and excessive manipulation that may damage vessels.

Context

GovInfoNow's records show Microvention Inc. has 15 other recalls on record since 2017-08-07. Recent brand recalls include a 2026-04-30 recall for Bobby Balloon Guide Catheters due to balloon pinhole defects and a 2026-04-06 recall for LVIS Stent Devices regarding vessel injury risks from manufacturing errors. Other recent recalls of similar products include an Olympus HX-400U-30 Ligating Device recall on 2025-10-30 and a Boston Scientific ENROUTE Transcarotid Neuroprotection System recall on 2026-07-09.

How to tell if you have this one

Check for serial numbers WDC-2/00842429103128, 0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345 in China and Japan. The product is part of the WEB Aneurysm Embolization System.

  • Brand
    Microvention
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Product Serial Numbers

    Verify if your WEB Detachment Controller has the listed serial numbers from the recall notice.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 17 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: nationwideWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: nationwideCONebraska: named in the noticeNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: nationwideARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: named in the noticeLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Florida, Illinois, Indiana, Kansas, Louisiana, Michigan, Minnesota, Nebraska, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Texas, Utah, Virginia.

Recall timeline

FDA published this recall 49 days after the firm started it.

1. Firm started the recall: Sep 21, 202212. FDA classified it Class II: Nov 2, 202223. Published in FDA’s Enforcement Report: Nov 9, 202232022
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0170-2023 (September 21, 2022) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0170-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA