Use cautionMedical device recall · FDA · Published September 21, 2022
Microvention WEB Detachment Controller Recall: Risk of Implant Misplacement
Microvention recalls WEB Detachment Controller (WDC-2) due to potential implant misplacement during aneurysm embolization. Affected units have specific serial numbers sold in China and Japan. Contact for remedy.
Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
Take action
Contact the company
📍Recalling firmMICROVENTION INC.35 Enterprise · Aliso Viejo, CA 92656
“Microvention WEB Detachment Controller Recall: Risk of Implant Misplacement.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/60032-microvention-web-detachment-controller-recall-risk-of-implant-misplacement. Based on FDA recall #Z-0170-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0170-2023%22&limit=1).
The FDA announced on 2022-09-21 that Microvention Inc. is recalling WEB Detachment Controllers due to a defect that may prevent proper detachment of implant devices. The agency classified the recall as Class II and ongoing.
How to tell if you have it
Brand: Microvention
Product: WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2
Component of WEB Aneurysm Embolization System
Units affected: 2018
The hazard
The detachment controller has an out-of-specification tunnel opening dimension that may prevent the implant device pusher from being fully inserted. This could result in the inability to properly detach the implant device, potentially causing delays and excessive manipulation that may damage vessels.
Context
GovInfoNow's records show Microvention Inc. has 15 other recalls on record since 2017-08-07. Recent brand recalls include a 2026-04-30 recall for Bobby Balloon Guide Catheters due to balloon pinhole defects and a 2026-04-06 recall for LVIS Stent Devices regarding vessel injury risks from manufacturing errors. Other recent recalls of similar products include an Olympus HX-400U-30 Ligating Device recall on 2025-10-30 and a Boston Scientific ENROUTE Transcarotid Neuroprotection System recall on 2026-07-09.
How to tell if you have this one
Check for serial numbers WDC-2/00842429103128, 0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345 in China and Japan. The product is part of the WEB Aneurysm Embolization System.
Brand
Microvention
Category
Medical device
Based on the official FDA remedy
What to do — step by step
Check Product Serial Numbers
Verify if your WEB Detachment Controller has the listed serial numbers from the recall notice.
Where it went and when
Where it was distributed
Distributed nationwide and abroad; the notice names 17 states.
Named in the notice
Covered by “nationwide”
Not listed
FDA distribution: “Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.” States read from FDA’s own wording; “not listed” means the notice does not name it.
Named in the notice: California, Florida, Illinois, Indiana, Kansas, Louisiana, Michigan, Minnesota, Nebraska, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Texas, Utah, Virginia.
Recall timeline
FDA published this recall 49 days after the firm started it.
1Firm started the recall
2FDA classified it Class II
3Published in FDA’s Enforcement Report
Dates from the agency record only; steps without a date are left out.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
Official source: FDA Recall #Z-0170-2023 (September 21, 2022) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0170-2023 to see it in context.
Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.