Corin recalls BIOLOX Delta ceramic heads with size labeling error that may cause incorrect fit. Affected units have size 36XL labeled as 32XL and vice versa.
The BIOLOX Delta Ceramic Head size 36XL is labelled as the size 32XL and vice-versa.
Corin is recalling BIOLOX Delta ceramic heads that have a size labeling error. This could lead to incorrect fitting, which might cause improper function or damage.
The size labels on the heads are swapped: size 36XL is labeled as 32XL and vice versa. This could lead to incorrect fitting, which might cause improper function or damage to the head or the system it's used with.
Users of Corin BIOLOX Delta ceramic heads with the specific model and lot number are affected. Those who have the heads labeled incorrectly are most at risk.
Check for the model number 036mm EX LONG +8mm, REF 104.3615 with UDI/DI (01) 0 5055196 96834 and Lot 503151. The size labels may show 36XL as 32XL and vice versa.
Verify the size labels on the head: size 36XL should be labeled as 36XL, not 32XL.
No, this recall is for a labeling error that could lead to incorrect fitting, but it does not pose a risk of serious injury or damage.
Check for the model number 036mm EX LONG +8mm, REF 104.3615 with UDI/DI (01) 0 5055196 96834 and Lot 503151.
Check the size labels to ensure they match the correct size. If they are swapped, contact Corin for guidance.
We’ve drafted a message you can send Corin to request your refund or repair — edit it as you like.
Subject: Recall Z-0727-2023 — Corin Corin Hello, I own a Corin that is covered by recall Z-0727-2023 from Corin. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.