Skip to content
Use caution Medical device recall · FDA · Published November 9, 2022

Siemens ARTISTE/ONCOR/PRIMUS Radiation Therapy Software Version 4.3.1 MR3 Recall: Wrong

Siemens recalls ARTISTE, ONCOR and PRIMUS radiation therapy systems with software version 4.3.1 MR3 due to risk of delivering radiation to the wrong body site during cancer treatment. Affected units have specific serial numbers and material numbers.

🇺🇸 FDA recall #Z-0472-2023Medical deviceSiemens ›
Take action

Contact the company

Recalling firmSiemens Medical Solutions USA, Inc40 Liberty Blvd · Malvern, PA 19355-1418
✎ GovInfoNow plain-language explainer

Siemens is recalling ARTISTE, ONCOR, and PRIMUS radiation therapy systems with software version 4.3.1 MR3. This software version may cause the system to select the wrong treatment site, potentially delivering radiation to the incorrect body area during cancer treatment. Read more at govinfonow.com/recall/59444-siemens-artiste-oncor-primus-radiation-therapy-software-version-4-3.

Source Summary by GovInfoNow · govinfonow.com/recall/59444-siemens-artiste-oncor-primus-radiation-therapy-software-version-4-3 · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0472-2023

Cite this page

“Siemens ARTISTE/ONCOR/PRIMUS Radiation Therapy Software Version 4.3.1 MR3 Recall: Wrong.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/59444-siemens-artiste-oncor-primus-radiation-therapy-software-version-4-3. Based on FDA recall #Z-0472-2023 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0472-2023%22&limit=1).

The FDA announced on November 9, 2022, that Siemens Medical Solutions USA, Inc. is recalling ARTISTE, ONCOR and PRIMUS linear accelerator systems with syngo RT Therapist software version 4.3.1 MR3 due to a hazard involving potential delivery of radiation doses to the wrong isocenter.

How to tell if you have it

  • Brand: Siemens
  • Product: ARTISTE, ONCOR and PRIMUS with syngo RT Therapist
  • Software version: 4.3.1 MR3
  • Material Number: 8162815
  • Serial Numbers: 10736, 10737, 10822, 10850, 10886, 10843, 10865, 10025

The hazard

According to the FDA, when selecting the site fraction group for treatment, users may select the wrong site for treatment and potentially deliver the dose to the wrong isocenter.

Context

GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Other recent recalls of similar products include Siemens ARTIS icono Ceiling Monitor Recall (June 8, 2026), Siemens ARTIS Icono Biplane Recall (May 20, 2026) and Siemens ARTIS Pheno VE30A/VE40A Recall (March 12, 2026). Other recent recalls of similar products include Elekta MOSAIQ Oncology System Software Recall (May 21, 2026), Elekta Leksell GammaPlan 11.1 Software Recall (December 12, 2025) and RayStation 2024B Software Recall (November 28, 2025).

How to tell if you have this one

Check for serial numbers 10736, 10737, 10822, 10850, 10886, 10843, 10865, or 10025 on the system. The software version must be 4.3.1 MR3 with material number 8162815.

  • Brand
    Siemens
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version is 4.3.1 MR3 and the material number is 8162815.

Where it went and when

Where it was distributed

Distributed nationwide; the notice names 7 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: named in the noticeNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: nationwideINOhio: named in the noticeOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: named in the noticeNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: named in the noticeSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: named in the noticeLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: nationwideTXFlorida: nationwideFL

FDA distribution: “US Nationwide distribution in the states of CA, IA, LA, NE, NV, OH, SC.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Iowa, Louisiana, Nebraska, Nevada, Ohio, South Carolina.

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Nov 9, 202212. FDA classified it Class II: Dec 9, 202223. Published in FDA’s Enforcement Report: Dec 21, 202232022
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0472-2023 (November 9, 2022) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0472-2023 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA