Cardiac Assist, Inc recalled LifeSPARC controllers and accessories via the FDA on Dec. 5, 2022, due to a software critical failure hazard. The agency announced the Class I recall of model numbers 5900-0000 and 5900-0001.
How to tell if you have it
- Brand: Cardiac Assist
- Product: SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY - 5900-0001
- Model No. 5900-0000 - UDI/DI 00814112020562
- Model No. 5900-0001 - UDI/DI 00814112020623
- Serial Numbers: C00115 through C00209 (full list in record)
- Quantity: 763 units
The hazard
The hazard involves a Critical Failure where the software freezes or crashes and the screen does not display data. This issue was first addressed in a recall initiated July 21, 2022.
Context
GovInfoNow's records show Cardiac Assist, Inc has 6 other recalls on record since 2019-11-19. Other recent recalls of similar products include HeartMate 3 System Controller Recall (2026-05-13) and Thoratec HeartMate 3 LVAS Kits Recall (2026-05-13).